12.4 Servicing Intervals
726 Titroprocessor, Short Introduction
44
44
12 GLP Functions
The 726 Titroprocessor supports the supervision of servicing intervals as
well as system validation and sensor tests.
12.4 Servicing Intervals
The maintenance of the 726 Titroprocessor should include a yearly service
check carried out by a specialist from your Metrohm supplier.
The service engineer will update the service data and enter an appropriate
service interval.
When the service interval has expired, the 726 Titroprocessor will display a
warning at start-up and advice the user to ask for a Metrohm Service.
The service data can be inspected on the GLP page. It is accessible by
pressing the <GLP> key on the keyboard.
ÚÄÄÄ GLP ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³ ³
³ Method Validations Date Interval Result ³
³ ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ ³
³ 1. ³
³ 2. ³
³ 3. ³
³ 4. ³
³ 5. ³
³ ³
³ ³
³ ³
³ ³
³ Instrument ident. : 01106 ³
³ Last servicing dated : 96-11-08 ³
³ done by : B. Miller ³
³ Program version : 5.726.0010 ³
³ ³
³ ³
³ ³
³ ³
³ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ Quit ÄÙ
ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿ ÚÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄ¿
³ ³ ³Delete ³ ³Define ³Extend.³Start ³ ³ ³Sample ³Sensors³
³ ³ ³method ³ ³method ³info ³valid. ³ ³ ³silo ³ ³
ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ ÀÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÙ
The service data will be updated by the service engineer.
12.5 Sensor Test
The 726 Titroprocessor offers a sensor test procedure that allows you to
determine whether your pH electrode is still working properly. For each sen-
sor, separate limits can be defined that must be fulfilled during the a sensor
test to ensure the suitability of the sensor for a specific (or common) appli-
cation.
The sensor test consists of an extended calibration with automatic control of
the stirrers Int.A or Int.B. The test can be carried out with any buffer set
containing the buffers pH 4 and pH 7. While we recommend to use
Metrohm buffers, other buffer sets can be used as well.