99. . Method Editor
Method Editor
726 Titroprocessor, Short Introduction
33
33
Use the TST( ) function for testing the presence of an endpoint.
R01=TST(EP1,0)*0,1*C01*157,45*100/SS
If no endpoint is found, the result R01 will not be invalid. The correct result
zero (0) will be indicated.
9.2.2 System Variables
All of the following system variables may be used for calculations.
Rxx
Result
RT1
Result of titer determination for validation
RT2
Theoretical result for validation
RE
Endpoint for syst. error calculation in validations
Xxx
Intermediate result
C##
Common Variable (of actual titration or measurement)
EP#
Endpoint volume
MNRxx
Mean value of results
MNXxx
Mean value of intermediate results
MNC##
Mean value of Common Variables
ID1
Sample identification 1
ID2
Sample identification 2
ID3
Sample identification 3
SS
Sample Size
DD
Determination Duration
MD
Measuring Duration; (of actual titration or measurement)
SD
Start Duration; for start conditions
IT
Initial temperature
ST
Start temperature
CT
Closing temperature
IM
Initial measured value
SM
Measured value at the start of a titration or measurement
CM
Measured value at the closing of a titration or measurement
SV
Dispensed start volume
CV
Dispensed volume at the closing of a titration
C
Concentration of reagent (available only with automatic rea-
gent check)
EM#
Endpoint measured value
EF#
Endpoint flag; endpoint found yes/no (1/0)
F#
Fixed endpoint volume
FM#
Fixed endpoint measured value
FF#
Fixed endpoint flag; fixed endpoint reached yes/no (1/0)
HNV#
HNP volume
HNP#
HNP measured value
AP
Asymmetry potential of calibrated sensor
SL
Slope of calibrated sensor
DRC
Signal drift in conditioning status
DOD
Dosing duration
ITD
Intermediate titration duration
$RN
Run number
$AS
Auto start counter
$CRF
Conditioning ready flag (Cond ok, yes/no (1/0)
$BC
Buffer counter