MI 3108 EurotestPV
Instrument operation
33
Warning:
Customized settings will be lost when this option is used!
If the batteries are removed for more than 1 minute the custom made settings will
be lost.
The default setup is listed below:
Instrument setting
Default value
Language
English
Contrast
As defined and stored by adjustment procedure
Z factor
0.80
RCD standards
EN 61008 / EN 61009
Commander
Enabled
Communication
RS232
Clamp settings
CLAMP 1
CLAMP 2
A1391, 40A
A1391, 40A
Solar settings
See chapter
4.4.10 Solar Settings
Test mode:
Function
Parameters / limit value
Sub-function
INSTALLATION:
EARTH RE
No limit
R ISO
L/N, L//E, N/E, L/L
No limit
Utest = 500 V
CONTINUITY
(r1,rN,r2, R2, R1+R2, R1+RN)
No limit
Z - LINE
Fuse type: none selected
VOLTAGE DROP
ΔU: 4.0 %
Z
REF
: 0.00 Ω
Z - LOOP
Fuse type: none selected
Zs
rcd
Fuse type: none selected
RCD
RCD t
Nominal differential current: I
N
=30 mA
RCD type: AC
non-delayed
Test current starting polarity:
(0
)
Limit contact voltage: 50 V
Current multiplier:
1
POWER:
CURRENT
C1
SCOPE
U,I
HARMONICS
U
I
U
h:1
ENERGY
I: 40A, U: 260A
SOLAR:
ISO PV
No limit
Utest = 500 V