MI 2883 Energy Master
Operating the instrument
73
UIF C. Pwr F. Pwr NF. Pwr
Flick Sym
H
iH Sig Temp
Harmonics alarms.
UIF C. Pwr F. Pwr NF. Pwr
Flick Sym H
iH
Sig Temp
Interharmonics alarms.
UIF C. Pwr F. Pwr NF. Pwr
Flick Sym H iH
Sig
Temp
Signalling alarms.
UIF C. Pwr F. Pwr NF. Pwr
Flick Sym H iH Sig
Temp
Temperature alarms.
Filters alarms according to phase on which
they occurred:
1
2 3 N 12 23 31 T
Shows only alarms on phase L1.
1
2
3 N 12 23 31 T
Shows only alarms on phase L2.
1 2
3
N 12 23 31 T
Shows only alarms on phase L3.
1 2 3
N
12 23 31 T
Shows only alarms on neutral channel.
1 2 3 N
12
23 31 T
Shows only alarms on phases L12.
1 2 3 N 12
23
31 T
Shows only alarms on phases L23.
1 2 3 N 12 23
31
T
Shows only alarms on phases L31.
1 2 3 N 12 23 31
T
Shows only alarms on channels which are not
channel dependent
1 2 3 N 12 23 31 T
Shows all alarms.
Selects an alarm.
Returns
to the “RECORDERS” submenu.
3.17 Rapid voltage changes (RVC) table
In this table captured RVC events are shown. Events appear in the table after finish,
when voltage is in the steady state. RVC events are measured and represented
according to IEC 61000-4-30. See 5.1.14 for details.
Figure 3.51: RVC Events table group view screen
F3