MI 2883 Energy Master
Operating the instrument
94
instrument lock prevents unauthorized changing of instrument working mode, it does not
prevent non-destructive operations as displaying current measurement values or trends.
User locks the instrument by entering secret lock code in the Lock/Unlock screen.
Figure 3.67: Lock/Unlock screen
Table 3.90: Description of Lock/Unlock screen
Pin
Four digit numeric code used for Locking/Unlocking the
instrument.
Press ENTER key for changing the Pin code. “Enter PIN”
window will appear on screen.
Note:
Pin code is hidden (****), if the instrument is locked.
Lock
The following options for locking the instrument are
available:
Disabled
Enabled
Table 3.91: Keys in Lock/Unlock screen
Selects parameter to be modified.
Change value of the selected digit in Enter pin window.
Selects digit in Enter pin window.
Locks the instrument.
Opens Enter pin window for unlocking.
Opens Enter pin window for pin modification.
Accepts new pin.
Unlocks the instrument (if pin code is correct).
Returns
to the “GENERAL SETUP” submenu.
Following table shows how locking impacts instrument functionality.
Table 3.92: Locked instrument functionality
MEASUREMENTS
Allowed access.
Waveform snapshot functionality is blocked.
RECORDERS
No access.
MEASUREMENT SETUP
No access.
ENTER