61
ATMOS 41W
EC basin
EC sensors
Figure 35 EC basin and EC sensor
d. Clean out any residue inthe basin including the back of the downspout screen.
e. Clean the screws using a mild detergent such as liquid dish soap and a nonabrasive
sponge or cloth.
NOTE: Avoid detergents that contain lotions or moisturizers.
f. Rinse the sensor and screws thoroughly with tap of deionized water.
g. Reattach the rain funnel (
Section 3.2.2.3
)
Contact
for a replacement Teflon screen.
CAUTION
DO NOT immerse the sensor in water. Water may corrode and ruin metal parts inside the sensors.
DO NOT touch the temperature sensor needle (
Figure 20
).
DO NOT paint the body or any part of the ATMOS 41W.
The temperature sensor needle lead wires are very delicate and can be easily damaged.
Avoid more than light pressure on the sonic transducers (
Figure 30
)
Be sure the instrument is level after cleaning