20
BGA-3500 Series Manual 5/01
6. Problem: The component slide and / or prism assembly seem to be “locked up”
or restricted.
Probable Cause:
• The transportation screws are not fully loosened, resulting in the inability of the component slide and / or
prism assembly to move freely.
Solution:
• Loosen transportation screws, which are located on the placement arm underneath the prism assembly
and on the top of the component slide.
Probable Cause:
• The locking arm has moved during transportation to the wrong side of the lock pin.
Probable Solution:
• Remove the side cover from the placement arm and lift arm into the correct position.
7. Problem: The illumination of the component greatly outweighs the illumination of the
PCB so that an image of the component leads is not obtainable (or vise versa).
Probable Cause:
• The lighting is not adjusted correctly.
Possible Solution:
• Adjust the component illumination and PCB illumination knobs so that both images are clearly visible.
8. Problem: You are unable to slide the prism in and out of the placement arm assembly.
Probable Solution:
• The locking arm for the Z-axis control knob is on the wrong side of the prism slider pin. Remove the side
cover and move the locking lever into the correct position.
MICRO OVEN:
1. Problem: When running the BGA-S1 Software, there is an error stating that communication cannot be
established between the PC and BGA System.
Check:
• Has the BGA/CSP-3500 Control Unit been turned on?
• Is the RS232 Cable securely connected between the BGA/CSP-3500 Series Control Unit and the PC?
• Has the correct communication port been selected?
Probable Solution:
• Turn the BGA/CSP-3500 Control Unit on.
• Ensure that a firm connection has been established between the BGA/CSP-3500 Control Unit and
the
PC.
• When running the BGA Software, select “Auto Detect” for communication port.
2. Problem: The Micro Oven nozzle is not mounting easily into the BGA-3528 Micro Oven Head Assembly.
Probable Cause:
• The nozzle lock / release lever has not been rotated fully counter-clockwise.
• The notch or oval hole on the Nozzle or BGA-NA Nozzle Adapter has not been indexed correctly.