The measuring-equipment SHAFT (ROHRTEST SP04) possesses the admission
of the LGA Würzburg for the test of fusible-ness-separators of the Prüfnorm
DIN in 1999-100.
We like to send you a copy of the certificate on demand.
Technical parameters
Connection:
over connection-cables at reason-appliance ROHRTEST 4,
Supply with protection-small-tension of the reason-appliance,
maximum cable-length 80m
Denseness-
DIN EN 1610, method "W"
tests: DIN 1999-100
DIN 4040-100
DIN 4261-1
DIN EN 12566-1
Special-tests (free parameters)
Measuring-area:
Max. 50 mm level-alteration
entspr. Max. 39 l loss of DN 1000
entspr. Max. 25 l loss of DN 800
Dissolution:
0.01 mm level-alteration
Precision:
+/-0.2 mm with expired or not existing calibration
+/-0.1 mm with valid calibration, See section 3.7
+/-0.03 mm with parallel temperature measurement and valid
calibration, See section 3.7
Please heed:
The stated measuring-precision is gained by the measuring
instrument under unfavorable conditions only 15 min after positioning
and switching on in the span required for this stabilization-process
being main from the temperature-difference between the place of the
storage as well as the transportation and the place of the use
dependent.
Delivery capacity: Measuring-equipment "shaft" with tripod and level unit,
extension-tubes 0,5 m and 1,0 m, Calibration report