Chapter 1. Overview
Page 2
FROGSCAN USER’S MANUAL
09/07/2018 Vs 003
1.3 Specifications
Environmental
Operating temperature: +10° C to +40° C
Storage temperature: -30° C to +70° C
Humidity: 95% non-condensing
Operational
Computers vary widely in their characteristics and performance. The exact pulse retrieval rate depends
on many factors including the speed of the processor, the speed of the bus, the speed of the memory,
and the graphics card. Typically, FROGscan runs from about 1 to 4 Hz on modern Core Duo computers
running a Windows 7 or later operating system.
Instrument Specifications
Parameter
Specification
Input Pulse Wavelength Range
450 nm - >2000 nm
Pulse Length Range
< 15 fs - 12 ps
Temporal Range
30 ps
Temporal Resolution
2 fs or better
Delay Increment
1 fs
Spectral Resolution
0.20 nm - 1 nm
Spectral Range
100 nm - 600 nm
Pulse Complexity
TBWP > 50
Intensity Accuracy
2%
Phase Accuracy
0.01 radians
Real-time Sensitivity (IpeakIave)
4 W
2
Averaged Sensitivity (IpeakIave)
< 0.1 W
2
Input Beam Size
2 - 8mm collimated
Nominal Polarization
Horizontal (vertical by rotating crystal)
Acquisition Speed
> 1 Hz 64 x 64 grid
Spectra required for measurement
Number in grid
FROGscan instruments are calibrated by the factory. Wavelength calibration is stored in the
spectrometer, and the temporal calibration is determined by the servo delay line. Typically, the delay line
is about 1 fs/step plus or minus a few percent. If you suspect that the temporal calibration is incorrect,
please contact the factory.