Mesa Labs Smart-Well Model 1710 Operation Manual - 9153 Rev E
13
6.
Using the touch screen display, complete the
BI Insert Wizard
by entering the
BI Lot Number
,
User ID
,
Sterilizer Number
, and
Cycle ID
.
7.
Upon completion of the
BI Insert Wizard
, the corresponding cavity’s
Query Menu
will be
displayed for confirmation of data entered. If the information is correct, press the
Back
button
to return to the
Main Menu
. If the information is not correct, press the
Edit
button to re-
launch the
BI Insert Wizard
.
Note: data cannot be modified after test completion.
8.
During test, an hourglass icon will appear on the
Main Menu
button corresponding to the
incubation cavity. The “sand” in the hourglass will move from the top of the icon to the bottom
as the test progresses. Pressing the button will bring up the
Query Menu
for the cavity, where
detailed test information is available and can be printed.
Test in progress
9.
Within the first 10 minutes of the incubation period, a BI may be removed for up to 10 seconds.
While the BI is removed, the LED in front of the incubation cavity will blink amber and a
countdown will appear on the
Main Menu
.
Note: If the BI is not re-inserted within 10 seconds, the test
will be canceled and cannot be resumed.
A BI in test for longer than 10 minutes should not be
removed or rotated in the test cavity. An initial reading of
the BI is stored for use in determining growth. For best
results, the BI should remain in the test cavity.
Note: the
BI Insert Wizard
must be completed prior to
obtaining test results. If the
BI Insert Wizard
is not
completed and left unattended, an audible beep will sound
periodically as a reminder.