403321001_001 2016-08-31
3.
Proximal Thermocouple (TC) on the SpineSTAR Ablation Instrument reaches 50ºC.
The RF Cycle Time can be reset to the pre-set value by pressing and releasing the Reset Cycle
Button.
The RF Total Time, number of Cycles, and RF Cycle Time can be reset together by pressing and
holding the Reset All button for 2 seconds.
Completion of Procedure
Turn off MetaSTAR RF Generator using the power button. Unplug the controller from the electrical
outlet (Mains) and remove the power cord from the back of the controller.
10.
PREVENTIVE MAINTENANCE, TROUBLESHOOTING AND REPAIR:
10.1.
Adjusting Volume
The MetaSTAR RF Generator has an adjustable volume control (Section 8.2 item 16) on the back of
the unit.
Twisting the adjustor clockwise will increase the volume.
10.2.
Maintaining Device Effectiveness
In the event of a blown fuse, only 5x20mm 3.15A/250VAC Type “T” (slow blow) fuses should be used
as replacements. Turn power off and disconnect the power cord from the electrical outlet. Remove
fuse holder by opening the Power Entry Module’s Fuse Drawer (Section 8.2 item 12) on the back of
the MetaSTAR RF Generator. Replace both fuses.
Besides the fuses there are no other user serviceable parts. No servicing or preventive maintenance
is required for this device. For servicing, replacement or repair, return cleaned unit (9.3) to
manufacturer.
11.
CUSTOMER SERVICE AND TECHNICAL SUPPORT, WARRANTY:
11.1.
Customer Service
Contact MERIT’s Customer Service for any customer or technical support.
Call
+1-800-356-3748
11.2.
Limited Warranty
Merit Medical Systems, Inc. (“MERIT”) warrants to the original purchaser of the MetaSTAR RF
Generator (“MetaSTAR RF Generator”) that the MetaSTAR RF Generator will be free of defects in
material and workmanship for a period of one year after the date of purchase when used as intended
under normal operating conditions and in conformance with their instructions for use and
maintenance instructions. The sole obligation of MERIT under this warranty shall be limited to the
replacement of the MetaSTAR RF Generator or equivalent models, at no charge to the original
purchaser and at the option of MERIT, during the warranty period but only if examination reveals to
the satisfaction of MERIT that the MetaSTAR RF Generator contains a defect in material or
workmanship.
This warranty is made in lieu of all other warranties, expressed, implied or statutory, including, but
not limited to the implied warranties of merchantability and fitness for a particular purpose and all
other obligations and liabilities on the part of MERIT. MERIT neither assumes nor authorizes any
other person to assume for it any other liability in connection with the sale of the MetaSTAR RF
Generator. This warranty shall not apply to MetaSTAR RF Generator or any other part thereof that
has been subject to accident, negligence, alteration, abuse, or misuse. MERIT makes no warranty
whatsoever with regard to accessories or parts used in conjunction with the MetaSTAR RF Generator
and not supplied and manufactured by MERIT.
The term “original purchaser,” as used in the warranty means that person or organization and its
employees, if applicable, to whom the MetaSTAR RF Generator was sold by MERIT. This warranty
may not be assigned or transferred in any manner.
In order to make a warranty claim, the original customer must obtain a Return Good Authorization
Number (“RGA Number”) and return instructions from MERIT’s Customer Service Department or
authorized representative. The allegedly defective MetaSTAR RF Generator must be returned,
cleaned, repackaged, clearly marked with the RGA number on all packaging, and adequately insured,
to the following address: Merit Medical Systems, Inc., 1600 West Merit Parkway, South Jordan, Utah
84095 U.S.A. A written statement must also be included that reasonably describes the defect(s) with
the MetaSTAR RF Generator and indicates the appropriate return address to ship the replaced
MetaSTAR RF Generator, as well as a copy of a receipt showing proof of purchase.