background image

 

403321001_001 2016-08-31 

 

   

Disclaimer 

Merit Medical Systems, Inc. reserves the right to change its products at any time in order to incorporate 
the most recent technological developments. This guide is subject to change without notice. 
Although this guide has been prepared with every regard to insure the accuracy of its contents, Merit 
Medical  Systems,  Inc.  assumes  no  liability  for  any  damages  incurred  from  the  application  of  this 
information. The recommendations are designed to serve as a general guideline. They are not intended 
to supersede institutional protocols or professional clinical judgment concerning patient care. 

Table of Contents 

 

Section Page 

1.

 

BRIEF DEVICE DESCRIPTION: ................................................................................................. 3

 

2.

 

INDICATIONS FOR USE: ........................................................................................................... 3

 

3.

 

CONTRAINDICATIONS: ............................................................................................................. 3

 

4.

 

WARNINGS: ............................................................................................................................... 3

 

5.

 

PRECAUTIONS: ......................................................................................................................... 4

 

6.

 

ENVIRONMENTAL PROTECTION: ............................................................................................ 4

 

7.

 

HOW SUPPLIED: ....................................................................................................................... 4

 

7.1.

 

Accessories: ....................................................................................................................... 4

 

7.2.

 

Single Use Devices 

– See table below for other devices required for Tumor Ablation. .......... 4

 

8.

 

INSTRUCTIONS FOR USE: ....................................................................................................... 5

 

8.1.

 

Set-Up .................................................................................. 

Error! Bookmark not defined.

 

8.2.

 

MetaSTAR RF Generator Nomenclature ............................... 

Error! Bookmark not defined.

 

8.3.

 

Initiation .............................................................................................................................. 7

 

8.4.

 

Setting Ablation Cycle Time ................................................................................................ 7

 

8.5.

 

Operation ............................................................................................................................ 7

 

9.

 

PREVENTIVE MAINTENANCE, TROUBLESHOOTING AND REPAIR: ....................................... 9

 

9.1.

 

Adjusting Volume ................................................................................................................ 9

 

9.2.

 

Maintaining Device Effectiveness ........................................................................................ 9

 

9.3.

 

Cleaning the MetaSTAR RF Generator.................................. 

Error! Bookmark not defined.

 

10.

 

CUSTOMER SERVICE AND TECHNICAL SUPPORT, WARRANTY: ......................................... 9

 

10.1.

 

Customer Service................................................................................................................ 9

 

10.2.

 

Limited Warranty ................................................................................................................. 9

 

11.

 

DISTRIBUTORS / AUTHORIZED REPRESENTATIVES: .......................................................... 11

 

11.1.

 

North America ....................................................................... 

Error! Bookmark not defined.

 

11.2.

 

Europe   ................................................................................ 

Error! Bookmark not defined.

 

APPENDIX A

 

PRODUCT SPECIFICATIONS ..................................................................................... 11

 

APPENDIX B

 

ELECTROMAGNETIC ENVIRONMENT GUIDANCE .................................................... 12

 

APPENDIX C

 

ABBREVIATIONS ........................................................................................................ 12

 

APPENDIX D

 

CONFORMANCE TO STANDARDS ............................................................................. 13

 

APPENDIX E

 

COMPATIBLE POWER CORDS .................................................................................. 13

 

APPENDIX F

 

TONES ........................................................................................................................ 13

 

APPENDIX G

 

INDICATORS / ERROR CODES AND TROUBLESHOOTING ...................................... 14

 

 

 

Содержание MetaSTAR 3195

Страница 1: ...West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Business Park West Galway Ireland 31...

Страница 2: ...PPLIED 4 7 1 Accessories 4 7 2 Single Use Devices See table below for other devices required for Tumor Ablation 4 8 INSTRUCTIONS FOR USE 5 8 1 Set Up Error Bookmark not defined 8 2 MetaSTAR RF Generat...

Страница 3: ...FU for the applicable device s used during the procedure Warning Hazardous electrical output This equipment is for use only by qualified medical personnel trained in the use of the MetaSTAR RF Generat...

Страница 4: ...istatic sheeting is recommended Caution Cables to the surgical electrodes are recommended to be positioned such that contact with patient or other leads is avoided Caution Do not continue to deliver R...

Страница 5: ...SpineSTAR Ablation Instrument Hand Switch Cable andAE Cable are available as backup prior to use 8 3 MetaSTAR RF Generator Nomenclature Symbols ON power Type BF Applied Part Volume Control OFF power W...

Страница 6: ...LOW LED 4 RF ON BLUE LED 5 POWER ON GREEN LED 6 Soft Key Button 1 7 Soft Key Button 2 8 Soft Key Button 3 9 Soft Key Button 4 10 Hand Switch Cable Socket 11 AE Cable Socket 12 Fuse Drawer 13 ON OFF Ma...

Страница 7: ...be connected to the AE Cable socket on the MetaSTAR RF Generator using the AE Cable inspect the AE Cable for damage If the SpineSTAR Ablation Instrument is connected properly Distal and Proximal tempe...

Страница 8: ...TAR Ablation Instrument is properly placed 1 Both thermocouple temperatures are displayed in amber 30 C Note temperatures are displayed in blue when temperature is 30 C 2 Electrode is extended to the...

Страница 9: ...instructions for use and maintenance instructions The sole obligation of MERIT under this warranty shall be limited to the replacement of the MetaSTAR RF Generator or equivalent models at no charge t...

Страница 10: ...erit Medical Systems Inc 1600 West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Busines...

Страница 11: ...operation Enclosure IP 21 III Operating Conditions Temperature 59 F to 104 F 15 C to 40 C Relative Humidity 30 to 75 non condensing Atmospheric Pressure 878 to 1082cmH2O 86 to 106kPa Whenever the tem...

Страница 12: ...Instrument are intended for use in the electromagnetic environment specified below The customer or the user of the System should assure that it is used in such an environment Immunity test IEC 60601...

Страница 13: ...ntific and Medical ISM RF Equipment MDD 93 42 EEC Medical Device Directive Council Directive Concerning Medical Devices APPENDIX E COMPATIBLE POWER CORDS Continental Europe 2 0m 10A Inter Power PN 862...

Страница 14: ...in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the...

Страница 15: ...e 39 F Watchdog Fault Turn Power off Turn Power back on If error persists contact MERIT Customer Service 41 53 W RAM Power On Self Test Turn Power off Turn Power back on If error persists contact MERI...

Отзывы: