background image

 

403321001_001 2016-08-31 

 

   

9. 

INSTRUCTIONS FOR USE: 

9.1. 

Initiation 

Once the device 

Preparation

 (8.0) is complete and power is turned on using the 

ON/OFF Main Switch

 

at the back of the MetaSTAR RF Generator, the system performs a self-check. All LEDs and pixels 
on the graphics display will be on and the speaker will sound for a few seconds upon initiation (Power-
On Tone). In case an error is detected during Self-Test the Fault or Warning LED will turn on and the 
Fault or Warning Tone will sound. 

NOTE

: The Hand Switch Cable can be inserted in the Hand Switch Cable Socket prior to powering 

up the MetaSTAR RF Generator.  
After self-check is performed, the software versions are displayed. 

9.2. 

Setting Ablation Cycle Time 

Upon completion of the self-check the generator will go to the set Ablation Cycle Time Screen.  In 
the Ablation Cycle Time Screen pressing the Up and Down Arrow Buttons will increase or decrease 
the RF Cycle time.  Once the desired time is selected, pressing the Confirm Button will enter the time 
and  advance  to  the  Main  Procedure  Screen.  During  operation  (below),  the  user  can  re-set  the 
Ablation Cycle Time by pressing and holding the Reset Cycle Button for 2seconds. 

9.3. 

Operation 
SpineSTAR Ablation Instrument and AE Cable 

“Connect Device” will flash until the SpineSTAR Ablation Instrument is connected to the MetaSTAR 
RF Generator. The SpineSTAR Ablation Instrument should be connected to the AE Cable socket on 
the  MetaSTAR  RF  Generator  using  the  AE  Cable  (inspect  the  AE  Cable  for  damage).  If  the 
SpineSTAR Ablation Instrument is connected properly, Distal and Proximal temperature readings will 
display. 

Hand Switch Cable 

The Hand Switch Cable should be inspected and inserted into the Hand Switch Cable socket on the 
MetaSTAR RF Generator. 
The MetaSTAR RF Generator will display the following information: 
1. 

RF Cycle: Displayed at the top left portion of the Generator Display in minutes:seconds. (00:00) 

2. 

Cycles: Displayed at the top center portion of the Generator Display in number of cycles (00) 
performed. 

3. 

RF Total: Displayed at the top right portion of the Generator display in minutes:seconds (00:00) format. 
This value represents the cumulative time RF power has been delivered. 

4. 

Distal TC:  Displayed  in  (ºC): Temperature  of  the  Distal Thermocouple  of  the  SpineSTAR  Ablation 
Instrument. 

5. 

Proximal TC: Displayed in (ºC): Temperature of the Proximal of the SpineSTAR Ablation Instrument. 

6. 

Ω (Impedance): Displayed as a box filling from bottom to top on bottom left of Generator’s display. 

7. 

Power: Displayed in (Watts): Ablation RF output power of the MetaSTAR RF Generator. 

RF Power Level 

This generator has four power settings, 3w, 5w, 7.5w and 10w. 
The selected output power should be as low as possible for the intended purpose. Certain devices 
or accessories may present SAFETY HAZARDS (See WARNINGS and PRECAUTIONS above). Set 
the power level to 5W and modify during procedure as required to create thermal distribution for the 
major and minor diameter as displayed in 

Figure 1

 

 

Содержание MetaSTAR 3195

Страница 1: ...West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Business Park West Galway Ireland 31...

Страница 2: ...PPLIED 4 7 1 Accessories 4 7 2 Single Use Devices See table below for other devices required for Tumor Ablation 4 8 INSTRUCTIONS FOR USE 5 8 1 Set Up Error Bookmark not defined 8 2 MetaSTAR RF Generat...

Страница 3: ...FU for the applicable device s used during the procedure Warning Hazardous electrical output This equipment is for use only by qualified medical personnel trained in the use of the MetaSTAR RF Generat...

Страница 4: ...istatic sheeting is recommended Caution Cables to the surgical electrodes are recommended to be positioned such that contact with patient or other leads is avoided Caution Do not continue to deliver R...

Страница 5: ...SpineSTAR Ablation Instrument Hand Switch Cable andAE Cable are available as backup prior to use 8 3 MetaSTAR RF Generator Nomenclature Symbols ON power Type BF Applied Part Volume Control OFF power W...

Страница 6: ...LOW LED 4 RF ON BLUE LED 5 POWER ON GREEN LED 6 Soft Key Button 1 7 Soft Key Button 2 8 Soft Key Button 3 9 Soft Key Button 4 10 Hand Switch Cable Socket 11 AE Cable Socket 12 Fuse Drawer 13 ON OFF Ma...

Страница 7: ...be connected to the AE Cable socket on the MetaSTAR RF Generator using the AE Cable inspect the AE Cable for damage If the SpineSTAR Ablation Instrument is connected properly Distal and Proximal tempe...

Страница 8: ...TAR Ablation Instrument is properly placed 1 Both thermocouple temperatures are displayed in amber 30 C Note temperatures are displayed in blue when temperature is 30 C 2 Electrode is extended to the...

Страница 9: ...instructions for use and maintenance instructions The sole obligation of MERIT under this warranty shall be limited to the replacement of the MetaSTAR RF Generator or equivalent models at no charge t...

Страница 10: ...erit Medical Systems Inc 1600 West Merit Parkway South Jordan Utah 84095 U S A 1 801 253 1600 U S A Customer Service 1 800 356 3748 Authorized Representative Merit Medical Ireland Ltd Parkmore Busines...

Страница 11: ...operation Enclosure IP 21 III Operating Conditions Temperature 59 F to 104 F 15 C to 40 C Relative Humidity 30 to 75 non condensing Atmospheric Pressure 878 to 1082cmH2O 86 to 106kPa Whenever the tem...

Страница 12: ...Instrument are intended for use in the electromagnetic environment specified below The customer or the user of the System should assure that it is used in such an environment Immunity test IEC 60601...

Страница 13: ...ntific and Medical ISM RF Equipment MDD 93 42 EEC Medical Device Directive Council Directive Concerning Medical Devices APPENDIX E COMPATIBLE POWER CORDS Continental Europe 2 0m 10A Inter Power PN 862...

Страница 14: ...in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the...

Страница 15: ...e 39 F Watchdog Fault Turn Power off Turn Power back on If error persists contact MERIT Customer Service 41 53 W RAM Power On Self Test Turn Power off Turn Power back on If error persists contact MERI...

Отзывы: