6.4.3.8 Detection Flags
The CPC4000 Industrial Pressure Controller is equipped with three detection flags that can be enabled
or disabled by the user as needed. These detection flags appear in the Control Settings App. The pri-
mary purpose of these detection flags is to protect the instrument and to ensure desired operation. Each
of these flags can be turned “On” or “Off” as per user’s need. Figure 6.4.3.8 shows the detection flag set-
tings.
• Supply Detection Enable: If turned “On”, this flag allows the user to check for sufficient supply pres-
sure at the Supply Port of the instrument. In case the supply pressure is less than over 10% of the
control setpoint, an error is reported which can be seen in the Troubleshoot App by clicking the error
symbol [ ]. This flag is turned “Off” by default.
• Burst Detection Enable: If turned “On”, this flag allows the user to protect the instrument against sud-
den bursts in the pressure at the Measure/ Control Port. In case a burst is detected, an error is report-
ed which can be seen in the Troubleshoot App. This flag is turned “On” by default.
• Measure Regulation Enable: If turned “On”, this flag allows pressure to be controlled while the instru-
ment is in “Measure” mode to prevent pressure leaks in the system over time. This will turn on the in-
ternal pressure regulator periodically to control pressure in a certain range of the measured pressure
value. This flag is turned “Off” by default.