90
Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
:INLimits?
Returns current pressure reading (float), stable
flag (0 not stable, 1 stable)
:INLimits
:TIME?
Returns the stable delay in seconds
:TIME<f>
Set the stable delay in seconds
RANGe
Set range hold on the desired sensor. Text must
include quotes, maximum sensor range, native
units and pressure type. i.e. “1000.000psig”.
:RANGe?
Returns the active sensor range text, i.e.
"1000.000psig"
:RESolution?
Returns sensor resolution (integer 4 to 6)
:RESolution <n>
Set the sensor resolution
:SLEW?
Returns the rate in current units
SOURce[x]
[x]=x=1=Channel A
[:PRESsure]
[:LEVel]
[:IMMediate]
[:AMPLitude]?
Returns the setpoint in current units
[:AMPLitude] <f>
Sets the setpoint in current units
:VENT?
Returns 1 if in vent mode and pressure is not fully
vented (stable) yet, 0 otherwise. Proper usage:
put in vent mode, then query this value until it is 0
again to know the system is vented.
:VENT ON | OFF | 1 | 0 ON or 1 puts in vent mode , OFF or 0 puts in
Measure mode
:SLEW?
Returns the rate setpoint in current units
:SLEW <f>
Sets the rate setpoint in current units
:SLEW
:MODe?
Returns MAX when highspeed mode is active.
Returns LIN otherwise.
:MODe <CCC>
Set the control mode by putting in either High-
speed or Custom mode. Note: the overshoot flag
and control mode are not independent in Mensor
products.
:OVERshoot?
Returns 1 if the channel’s control behavior is >50.
Returns 0 if the control behavior <=50.
:OVERshoot ON | OFF | 1 | 0
ON or 1 puts in “overshoot allowed” mode, where
if the control behavior <90, it will be set to 100.
OFF or 0 puts in low overshoot mode, where if
the control behavior is >10, it will be set to 10.