70
Operating Instructions - CPC3050
High-Speed Pressure Controller
CPC3050
Listcal?
<sp>PRI,{sn},1,{mmddyy};SEC,{sn},
1,{mmddyy},{sn},1,{mmddyy},BAR,{
sn},{ mmddyy}<cr><lf>
Returns list of available transducers
including serial number and date of
calibration
Listrange?
PRI,1,min,max;SEC,1,min,max;Bar,m
in,max
Returns the ranges of the installed
transducers.
Listsensorinfo?
<sp>PRI,SN,PTYPE,ACCURACY,1,m
in,max,2,min,max;SEC,SN,PTYPE,A
CCURACY,1,min,max,2,min,max;<cr
><lf>
Returns the serial number, pressure
type, accuracy, and ranges of the in-
stalled sensors for the active channel
LowerLimit
Value inside primary xducer range in
current units.
Sets the lower control limit for the
instrument
LowerLimit?
<sp>n.nn<cr>
<lf>
Returns the lower control limit for the
instrument in current units.
Measure
None
Instrument placed in Measure Mode
Measure?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in mea-
sure. NO if otherwise
MediaDensity
NITROGEN or DRYAIR
Set the media type to either the gas
nitrogen or dry air.
MediaDensity?
<sp>ccccc<cr><lf>
Returns the media type
MediaTemp
Value
Set the media temperature in active
head pressure correction temperature
units.
MediaTemp?
<sp>+n.nn<cr><lf>
Returns the media temperature in the
active head pressure correction tem-
perature units, as either imperial (°F)
or metric (°C).
Mode
MEASURE, CONTROL, VENT
Sets the operation mode
Mode?
<sp>XXXXXX<cr><lf>
Returns the operation mode
Netmask
nnn.nnn.nnn.nnn
Sets the Ethernet network mask
Netmask?
<sp>nnn.nnn.nnn.nnn<cr><lf>
Gets the Ethernet network mask
OSversion?
<sp>n.n<cr><lf>
Returns the operating system version
Outform
1 to 7 – see table in Section 7.8
Sets the output format
Outform?
<sp>X<cr><lf>
Returns the output format – see table
in Section 7.8
Peakmax?
<sp>n.nn<cr>
<lf>
Returns the maximum pressure since
peakreset was sent.
Peakmin?
<sp>n.nn<cr>
<lf>
Returns the minimum pressure since
peakreset was sent.
Peakreset
None
Resets the peak values.
Port
nnnnnn
Sets the Ethernet port of the instru-
ment
Port?
<sp>nnnnn<cr><lf>
Returns the Ethernet port of the in-