6.4.10 Step Settings Application
The Step Settings Application allows the user to enter control setpoint value or write program
sequences as a factor of the percentage of the device under test operating pressure range or
as fixed pressure values (figure 6.4.10). The user can enter the minimum and maximum range
of the device under test (figure 6.4.10-A) and then select the desired percentage value(s) or
desired fixed pressure value(s) for setpoint(s). These ranges should be within the maximum
range of the primary transducer.. The user can also enter the control point up to a maximum of
5% over the range of the device under test limits by entering the overrange percentage (figure
6.4.10-B). CPC3050 automatically calculates the pressure value of the setpoint based on this
selection. The user can be directed to this application from the setpoint entry menu in the Home
App (section 6.4.1.2).
Figure 6.4.10 – Step Settings Application
Figure 6.4.10-A – Range Selection
Figure 6.4.10-B - Overrange Selection