Operating Instructions - CPC3050
99
High-Speed Pressure Controller
CPC3050
Wear safety goggles when using and/ or servicing the Automatic CPS to protect
eyes from flying particles and liquid splashes
Warning
!
Warning! The Automatic CPS should only be used to decontaminate the system
for liquid or particulate contaminents that are non reactive, non corrosive and
non explosive.
8.6.1 Automatic CPS Installation
The Automatic CPS can be mounted either in a wall/ rack mount configuration by the keyholes (Figure
8.6.1-A) on either or both side of the accessory or placed on a steady surface as a desktop accessory.
The DUT can be mounted either on top of the accessory for convinient test stand setup or on the back
of the accessory for connection with a pressure hose (Figure 8.6.1-B). The not connected port should be
sealed by the plug provided to ensure no leakage.
The pressure connection to the Measure/ Control port of the CPC3050 is established by a pressure tube
with 7/16” SAE adapters. The Automatic CPS is powered by the CPC3050 using a 3 meter long flexible
cable which connects to the Automatic CPS Backplate on the instrument.
[ 38.1mm ]
1.500in
[ 177.8mm ]
7.000in
[ 3.2mm ]
R.125in
[ 6.4mm ]
R.250in
MENSOR
AUTOMATIC CPS LABEL
Figure 8.6.1-A - Side view of Automatic CPS with mounting keyholes