8
D32802 | Date 11/2018
Description and function
2. Description and function
2.1 Description
The appliance can heat the interior up to 70 ºC and cool it down to 0 ºC. In addition to
the HPP standard equipment, the appliance features an exhaust fan in the back of the
chamber and 4 swivel-type LED light strips for adjusting the interior lighting in steps of
1 %. HPPlife appliances have more humidity. Therefore, use the central compressed air
dehumidification (see page 13).
2.2 Working range
The temperature-humidity diagram (Fig. 2) specifies at what range of temperature and
humidity a permanent, condensation-free operation is possible.
Caution:
Always observe the indicated humidity-temperature range. Otherwise, puddles may
form in the chamber and leak through the door seal.
0
10
20
30
40
50
60
70
80
90
100
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
Temperature °C
Humidity
% rH
Temperature
Fig. 2
Temperature-humidity working range of animal keeping chambers HPPlife