background image

D32802 | Date 11/2018  

11

Putting into operation

3.2  Connecting the appliance

3.2.1  Power and water

Connect the appliance to the power and water supply as described in the operating 

manual for the constant climate chamber HPP.

3.2.2  Compressed air

Warning! Only connect compressed air to the gas connec-

tion on the back of the appliance. Introducing other gases 

or substances can cause severe damage and considerable 

danger (explosion or poisoning).

Use a suitable pressure hose to connect the 

chamber to an external compressed air 

supply (max. 5 bar) at the rear of the 

appliance (Fig. 7). 

Only oil-free compressed air may be 

used.

See page 13 to see instructions on how 

to set compressed air dehumidification .

3.3  Switching on the appliance

Switch on the appliance by pressing the 

main switch on the front of the appliance 

( Fig.  8 ).

The appliance displays are in English by 

default when the appliance is switched 

on for the first time. You will find 

instructions as to how to change the 

language in the operating instructions of 

the constant climate chamber HPP.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen, 

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung 

getroffen hat. 

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

 

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Fig. 8 

 Switching on the appliance

Fig. 7 

 Compressed air connection (max. 

5 bar)

Fig. 7 

Fig. 7 

Содержание HPPlife

Страница 1: ...www memmert com www atmosafe net 100 ATMOSAFE MADE IN GERMANY SUPPLEMENT TO THE OPERATING MANUAL CONSTANT CLIMATE CHAMBER HPP HPPlife ANIMAL KEEPING CHAMBER HPPlife...

Страница 2: ...E mail service memmert com For service enquiries please always specify the appliance number given on the type plate Shipping address for repairs Shipping address for repairs Memmert GmbH Co KG Custome...

Страница 3: ...functions described in this manual refer to the latest firmware version Due to individual configurations and sizes illustrations in this manual may be slightly dif ferent from the actual appearance F...

Страница 4: ...ption 8 2 2 Working range 2 2 Working range 2 2 Working range 8 2 3 Technical data 9 2 4 Scope of delivery 2 4 Scope of delivery 2 4 Scope of delivery 9 3 Putting into operation 10 3 1 Assembling the...

Страница 5: ...2 Product safety and dangers The appliances described in this manual are technically sophisticated manufactured using high quality materials and subject to many hours of testing in the factory They r...

Страница 6: ...for working with experimental animals are complied with is responsible for ensuring that unauthorised persons have no access to the appli ance is responsible for ensuring that the maintenance plan is...

Страница 7: ...ctions and irregularities The appliance may only be used in a flawless condition If you as the operator notice ir regularities malfunctions or damage immediately take the appliance out of service and...

Страница 8: ...more humidity Therefore use the central compressed air dehumidification see page 13 2 2 Working range The temperature humidity diagram Fig 2 specifies at what range of temperature and humidity a perm...

Страница 9: ...6 5 115 V 50 60 Hz 13 max number of sliding grids max 3 illuminated max load per sliding grid kg 30 max load per appliance kg 200 Temperature Adjustment range 0 C to 70 C Working range with illuminati...

Страница 10: ...it 2 2 Insert the exhaust air unit with the pipe socket into the entry port in the back of the chamber Fig 4 and attach it to the chamber with the eight Torx screws included Fig 5 Make sure that the e...

Страница 11: ...ompressed air dehumidification 3 3 Switching on the appliance Switch on the appliance by pressing the main switch on the front of the appliance Fig 8 The appliance displays are in English by default w...

Страница 12: ...y adjustment Setpoint and actual temperature display adjustment 3 Fan speed display Setpoint and actual temperature display Fan speed display Setpoint and actual temperature display 4 Activation key f...

Страница 13: ...limate chamber HPP for the adjustment options of the timer the programme operation tempera ture and humidity monitoring the graphical representation and the menu functions 4 3 Setting compressed air d...

Страница 14: ...ctory is 0 2 bar This is enough to achieve a certified dehumidification of 10 humidity at 10 C under laboratory conditions You can adjust the operating pressure after testing the appliance under the o...

Страница 15: ...0 fan speed 20 LED lighting and 22 C for 30 minutes Day simulation 60 fan speed 60 LED lighting and 22 C for 12 hours 5 Night simulation 0 LED lighting and 22 C for 11 hours with unchanged fan speed o...

Страница 16: ...emmert Stra e 90 96 D 91186 B chenbach Tel 49 9122 925 0 Fax 49 9122 14585 E Mail sales memmert com facebook com memmert family Die Experten Plattform www atmosafe net Animal keeping chamber HPPlife D...

Отзывы:

Похожие инструкции для HPPlife