Operation and control
4.2 Adjustment options
You can set the following parameters after pressing the corresponding activation key (in
any sequence):
Temperature
Temperature
Adjustment range: 0 to 40 °C
Heating operation is indicated by the
symbol.
Cooling is indicated by the
symbol.
°C
22
.4
TEMP
°C
22
.0
Set
Humidity
Humidity
Adjustment range: 10 to 90 % rh
Humidification is indicated by the
symbol.
Dehumidification is indicated by the
symbol.
FEUCHTE
34
%
70
%rh
Set
rh
LED interior lighting
LED interior lighting
Adjustment range: 0 to 100 % in steps of 1%
LICHT
%
40
Fan speed
Fan speed
Adjustment range: 0 to 100 % in steps of 10%
Make sure that no draught forms in the interior due to the
settings.
LÜFTER
%
50
Please refer to the operating manual for the constant climate chamber HPP
for the adjustment options of the timer, the programme operation, tempera-
ture and humidity monitoring, the graphical representation and the menu
functions.
4.3 Setting compressed air dehumidifaction
The compressed air cannot be dehumidified with the ControlCOCKPIT. Use the pressure
controller (Fig. 12) instead. For best results when dehumidifying compressed air, set the
humidity range so that it fluctuates around the setpoint with the smallest amplitude pos-
sible. Use the ControlCOCKPIT (Display "
Graph"
) or the AtmoCONTROL software to track
the relative humidity in the appliance (Fig. 10 and Fig. 11).
Fig. 10
Incorrect operating pressure (high
humidity fluctuation)
Fig. 10
Fig. 10
Incorrect operating pressure (high
Incorrect operating pressure (high
Fig. 11
Correct operating pressure (small
humidity fluctuation)
Fig. 11
Fig. 11
Correct operating pressure (small
Correct operating pressure (small