5.6.5
Delete all stored logs ____________________________________________________________
In order to delete all cycle logs (e.g. in the event of the warning message "Protocol memory full", with the
option "Immed. output NO", selected (see Section 9.1), then after switching on the steam sterilizer proceed as
follows:
Operation
Display message
1. Hold down "+" button and also press "-",
Select "Function" menu, submenu "Last batch number"
2. Navigate with button "+" (or "-") until the display shows
submenu "Batch output"
3. Press button "Program", display shows menu "Batch output",
submenu "Output medium"
4. Press "+" (or "-") until the display shows "Delete all cycles".
5. Press the "Program" button to delete all logs
(or press "Start-Stop" to terminate).
6. Then press "Start-Stop" to return to "Function" menu...
7. ...and press "Start-Stop" again to return to the starting display
Func ti on:
La s t ba tc h num ber
Func ti on:
Ba t c h out put
Func ti on
Ba t c h out put
Al l c yc l e s
de le te
Al l oc a te d: 0
Fr ee : 40
1 4 :2 7 :12
- 0. 02 bar 25 °C
Ba t c h out put
O ut put me di um