MELA
tronic
23EN
18
Trays
Trays for objects which are to be sterilized must be perforated, in order to get safe sterilization results and to
allow condensation to run away. MELAG trays are recommended. If you use dishes or trays without
perforations, then the sterilization will not be safe and the steam sterilizer will not dry properly.
Closed sterilization containers
Closed sterilization containers must be perforated on two sides, in order to ensure that steam can penetrate
and condensate can run out. All MELAG sterilization containers meet these requirements.
If sterilization containers are stacked in the steam sterilizer, it is important to ensure that the perforations are
not blocked.
Transparent sterilization packaging
If you use transparent sterilization packaging, such as MELAfol, then the items should be stood vertically on
the tray. For the steam sterilizer the package holder (art. no. 00283) should be used. The packaging should
never be laid flat one on top of the other.
NOTICE
If seals split open during sterilization it may be necessary to increase the length of the
impulse on the sealing device or to use a double-seal.
Maximum loads
Do not exceed the maximum loads of the steam sterilizer.
Mixed loads
If different types of packaging are included in a load, then:
•
Instruments and sterilization containers should be at the bottom
•
Transparent and paper sterilization packaging should be at the top
Liquids
The steam sterilizer is not suitable for the sterilization of liquids!
Suitability for sterilization
Relevant information provided by manufacturers of instruments about sterilization should be strictly observed.
4.4 Capacity/Weight ____________________________________________________
Mount
Trays
Sterilization containers
mount “A”
art. no. 40248
package holder
art. no. 00283
max. 5 trays
art. no. 00230
4x 17K
2x 23M
1x 23G
4.5 Closing the door
____________________________________________________
PLEASE NOTE
Please close the door only when the steam sterilizer is switched on so that the door locking bolt goes back. It is
a precondition to close the door fully by turning the hand wheel.
The door is closed by putting the door latch over the door beam and turning the hand wheel until the door is
closed.