User Manual 

 

Steam sterilizer 

 

MELA

tronic

 

23EN

 

from software version 5.15 

 

 

 

 

 

 

Dear customer, 
 
We thank you for your confidence demonstrated by the purchase of this MELAG product. 
As an owner-run and operated family concern founded in 1951, we have a long history of successful 

specialization in hygiene products for practice-based use. Our focus on innovation, quality and the highest 

standards of operational reliability has established MELAG as the world’s leading manufacturer in the 

instrument treatment and hygiene field. 
You, our customer are justified in your demand for the best products, quality and reliability. Providing 

“competence in hygiene

” and 

“Quality – made in Germany”

, we guarantee that these demands will be met. 

Our certified quality management system is subject to close monitoring: one instrument to this end is our annual 

multi-day audit conducted in accordance with ISO 13485. This guarantees that all MELAG products are 

manufactured and tested in accordance with strict quality criteria. 
 

The MELAG management and team. 

 

 

 

EN 

 

 

 

Содержание MELAtronic 23EN

Страница 1: ...has established MELAG as the world s leading manufacturer in the instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Prov...

Страница 2: ...chnik GmbH Co KG Geneststra e 6 10 10829 Berlin Germany E Mail info melag com www melag com Original instructions Responsible for the contents MELAG Medizintechnik GmbH Co KG Subject to technical chan...

Страница 3: ...g instruments for sterilization 17 4 3 Loading the steam sterilizer 17 4 4 Capacity Weight 18 4 5 Closing the door 18 4 6 Program selection 19 4 7 Select automatic pre heating 20 4 8 Program start 20...

Страница 4: ...NS 40 9 1 Warning messages 40 9 2 Malfunction messages 42 9 3 Reaction to warnings error messages 44 9 4 No display on the screen 44 9 5 Door lock blocked 44 9 6 Leaks in the door 45 9 7 Too large fee...

Страница 5: ...the correct order Hazard avoidance Please read the safety information carefully before using the steam sterilizer About this manual Symbols Meaning Explanation Warning Health hazard Indicates a dange...

Страница 6: ...ockist or customer services using only original parts and following service instructions Documentation media computer CF card reader etc must be placed in such a way that they cannot come into contact...

Страница 7: ...ty to your sterilizer It represents a component of the product In affixing this CE mark the manufacturer declares that this medical product fulfils the basic requirements of the medical device directi...

Страница 8: ...ch you no longer require in the appropriate manner Comply with all relevant disposal specification in terms of possibly contaminated waste Packaging The packaging protects the device against transport...

Страница 9: ...r steam sterilization Failure to observe these provisions can result in damage or can compromise safety User Benefits Universal use The steam sterilizer sterilizes on the basis of the fractionated flo...

Страница 10: ...re time and pressure during a program run It monitors the parameters in terms of their threshold values during control and regulation and guarantees safe and successful sterilization If one or more pa...

Страница 11: ...MELAtronic 23EN 11 2 Device Description 2 1 Device views_______________________________________________________ Fig 1 Device views 1 0 1 0 9 1 2 1 1 1 3 8...

Страница 12: ...7 Door 21 Door grip 8 Back feet 22 Locking latch 9 Mains power cable 23 Tray 10 Fuses 2x 16 A FF 24 Mounting for trays 11 Connector for external condensate container 25 Door gasket 12 Serial data and...

Страница 13: ...chamber are not possible 2 3 5 One way Closed loop system Conductivity measurement _____________________________ The steam sterilizer operates in the standard installation mode as a circulatory flow...

Страница 14: ...__________________________________________________________ By activating the pre heating function the cold steam sterilizer chamber can be warmed up before sterilization or the temperature can be main...

Страница 15: ...26 you need at least a space of 30 cm or more above the steam sterilizer The electrical power supply should be a separate 220 240 V AC circuit with a 16 A fuse and optionally protected with a 30 mA re...

Страница 16: ...______________ To operate the steam sterilizer in the one way water system and to avoid discharge of steam during frequent sterilization work an external condensate container art no 00356 can be conne...

Страница 17: ...codes of practice in a disinfectant and or cleaning solution at the correct concentration for the correct length of time MELAG recommends the use of cleaning aids such as ultrasonic baths cleaning an...

Страница 18: ...during sterilization it may be necessary to increase the length of the impulse on the sealing device or to use a double seal Maximum loads Do not exceed the maximum loads of the steam sterilizer Mixe...

Страница 19: ...of unwrapped and solid instruments no hollows for rapid re use The maximum loads must not be exceeded Gentle Program at 121 C 1 bar sterilization time 20 5 min for the sterilization of all types of si...

Страница 20: ...on menu submenu Last batch number 2 Press or until the display shows the submenu autom Preheating 3 Press Program and the display shows the current option here Preheating YES 4 Pressing the Program bu...

Страница 21: ...ture and the time remaining 4 Pressure release After completion of the sterilization time the pressure is released and the steam generator emptied Pressure and temperature fall 6 Program end After ven...

Страница 22: ...play message 1 Press the Start Stop button To confirm press the Start Stop once again within 5 seconds If no confirmation is given then the program resumes normally 2 If confirmation is given then the...

Страница 23: ...ried after a successful sterilization cycle about 30 minutes when door opened about 5 cm and the preheating is activated Directly after sterilization there may still be residual condensation on the it...

Страница 24: ...yed as follows Operation Display message 1 Hold down button and also press select Function menu submenu Last batch number 2 Press or until the display shows the submenu Batch counter 3 Press Program t...

Страница 25: ...e g with MELAtrace MELAview Log printer MELAprint 42 44 MELAflash CF Card Printer MELAnet Box The installation and the operation of these devices is provided in Section 5 2 and the following From devi...

Страница 26: ...dium _______________________ 5 2 1 Connecting the log printer _________________________________________________________ In order to connect a log printer to the steam sterilizer then a printer cable s...

Страница 27: ...steam sterilizer Display shows 1 Line Time 2 Line Pressure of chamber Temperature inside chamber 2 Hold down button and also press button The Function menu with the submenu Last batch number is shown...

Страница 28: ...s Select Function menu submenu Last batch number 2 Press Program button Select Function menu submenu Batch output 3 Press Program button the display shows the menu Batch output submenu Output medium 4...

Страница 29: ...scribed in Section 5 2 2 under step 6 select MELAflash or MELAnet in the Function menu as output medium 5 5 No output medium __________________________________________________ In order to select the o...

Страница 30: ...ay message 1 Hold down button and also press button Select Function menu submenu Last batch number navigate with button or until the display shows submenu Batch output 2 Press Program button select Ba...

Страница 31: ...ory can be viewed as follows Operation Display message 1 Hold down button and also press button Select Function menu submenu Last batch number 2 navigate with button or until the display shows menu Fu...

Страница 32: ...tput menu submenu Output medium 3 Press or until the submenu Output all cycles appears on the display 4 Press Program button to start the output of all the stored logs up to 40 or to terminate press S...

Страница 33: ...Function menu submenu Last batch number 2 Navigate with button or until the display shows submenu Batch output 3 Press button Program display shows menu Batch output submenu Output medium 4 Press or u...

Страница 34: ...ation Steam entry 1 00 120 5 16 23 Press release 0 20 105 7 17 06 Heat up 2 05 134 3 22 08 Steriliz begin 2 05 134 3 22 08 Steriliz end 2 18 135 9 27 38 Press release 0 19 105 6 28 52 Drying end 0 19...

Страница 35: ...water vapour it is advisable to connect the external condensate container In case of using the closed loop system please observe the max sterilization cycles per work day see Section 2 3 5 page 13 6 2...

Страница 36: ...sterilizer operates with easily accessible filters The condition of these filters is very important for the proper functioning of the steam sterilizer Please service the filters as follows If you are...

Страница 37: ...they are made of stainless steel or which are made of normal steel but which have a damaged galvanic coating Often a single rusty instrument is enough to pass rust on to other instruments or to lead...

Страница 38: ...ccur and that may be the result of excessive wear on the door locking system If any of these difficulties occur inspection will be necessary by a MELAG Customer Service representative or by an authori...

Страница 39: ...ror reports can be made immediately If a program is completed without problems then on its completion there is an End message The print out contains a corresponding report The operator of the steam st...

Страница 40: ...ing occurs repeatedly please consult your stockist Warning message Cause Remedy Door not closed properly Turn the door grip down until contact is made display should then show Door closed This message...

Страница 41: ...stered and in the Batch output menu the option Immed output NO is selected The message is displayed when a program is started Pressing the Start Stop button again deleted the message and the program s...

Страница 42: ...for the heating up phases during air removal by sub atmospheric pulsing but also for achieving the required sterilization pressure Causes of this error may be any of the following Maximum loading amou...

Страница 43: ...eheating and inform your stockist The monitoring time for the pressure release in the outflow process for the fractionating was exceeded Check that the flow filter in the chamber at the bottom directl...

Страница 44: ...ere is no display Check 1 Is the power cable plugged into the mains 2 Is the mains supply OK if necessary check with another appliance 3 Exchange the two power fuses Fig 1 Pos 10 on the back of the st...

Страница 45: ...cycle is not condensed and returned to the water storage tank but is collected in an external condensate container In this case the consumption of water will depend on the program and the loading of...

Страница 46: ...ter one way 450 ml Program running time maximum load 34 min Sterilisation time 5 5 min Sterilization temperature range 4 0 K Quick Program S 2 1 bar 134 C Energy consumption min max 360 Wh Feed water...

Страница 47: ...inlays 00286 Handle for standard tray cassette 28895 Sterilization container 17K with 100 paper filters 01171 Sterilization container 23M with 100 paper filters 01231 Sterilization container 23G with...

Страница 48: ...t the steam sterilizer is not overloaded and that the load is arranged properly to ensure good drying The standard option autom Preheating serves to influence the program progress Any further program...

Страница 49: ...hamber test is located near the door The hottest point see in the sterilization chamber during the empty chamber test is located in the upper left corner of the tray rack at the rear end of the chambe...

Страница 50: ...MELAtronic 23EN 50 Fig 6 Prion Program 134 C and 2 1 bar Fig 7 Gentle Program 121 C and 1 1 bar Fig 8 Quick Program 134 C and 2 1 bar Pressure mbar rel Pressure mbar rel Pressure mbar rel...

Страница 51: ...Silicon oxide SiO2 1 mg l Iron 0 2 mg l Cadmium 0 005 mg l Lead 0 05 mg l Other heavy metals 0 1 mg l Chlorides Cl 2 mg l Phosphates P2O5 0 5 mg l Conductivity at 20 C 15 S cm pH degree of acidity 5...

Страница 52: ...nstable state even low level agitation can produce a large bubble within Demineralized water also called aqua dem water without the minerals usually found in normal spring or tap water is produced tho...

Страница 53: ...wrapped instruments sealed in a double layer of film or wrapped in film and placed in an additional container or a container wrapped in textiles Porous permeable for fluids and air e g textiles Porous...

Страница 54: ...devoid of all material in the technical sense volumes with a reduced gas pressure at least air pressure VDE Verband der Elektrotechnik Elektronik und Informationstechnik e V Alliance of the Electroni...

Страница 55: ...User Manual MELAtronic 23 EN...

Отзывы: