Chapter 5 – Logging
40
Delete saved logs
In order e.g. to suppress the warning text
Protocol memory full
for
preset option
Immediate output No
, the saved logs can be manually
deleted.
In the following example you can read how you can delete all saved logs.
Select the menu
Function
by simultaneous brief pressing of the
K
EYS
(+) and (-).
The display shows
Function: Last batch number.
.
Navigate with the K
EY
(+) or (-), up to the display
Function: Batch output
is shown and press the K
EY
(P).
Navigate with the K
EY
(+) or (-) until on the display
All cycles
delete
is shown.
Press the K
EY
(P) in order to delete all logs now.
In order to cancel the submenu without deleting, press the K
EY
(S).
In order to leave the menu after deletion, press the K
EY
(P). The
display shows
Function: Batch output
.
By again pressing the K
EY
(S), you completely leave the menu and
return to the starting position of the display.
Display protocol memory
With a connected and initialised printer or another output medium you can
verify how many logs are already saved in the log memory of the steam
sterilizer.
Proceed as follows:
Select the menu
Function
by simultaneous brief pressing of the
K
EYS
(+) and (-).
The display shows
Function: Last batch number.
.
Navigate with the K
EY
(+) or (-) until on the display
Function:
Batch output
is shown and press the K
EY
(P).
Navigate with the K
EY
(+) or (-), up to the display for the storage
occupancy (see left display figure) are shown.
In order to leave the menu again, press the K
EY
(S) twice.
Reading logs correctly
Header
The head of the program listing includes the general fundamental
information for the completed program such as for example current date,
the selected program, the daily batch number and the steam sterilizer
type.
Program step values
During the program the program sequence phases are recorded with the
accompanying values for steam pressure, temperature and time (relative
to the program start).
Summary
The summary states whether the program was successfully concluded.
Furthermore the values of the required sterilization time, the sterilization
temperature and the pressure including its maximum deviations are
displayed.
All cycles
delete
Allocated: 40