Chapter 3 – Initial Commissioning
20
Example 2 – MELAdem 47
(1)
(14)
(6)
(10)
(9)
(2)
(7)
(8)
(15)
(16)
(17)
(5)
(4)
(11)
(3)
(13)
(18)
(12)
Purified water from the reverse
osmosis unit MELAdem
47
Drain via waste water container
Fig. 11: Feed water supply from the reverse osmosis unit MELAdem
47
(1) Rear view Vacuklav 40 B+
(2) Safeguard combination, consisting of return flow inhibitor and pipe aerator according to EN 1717
(3) Y piece 3/4” (Art. No. 37300)
(4) Cooling water inflow fitting
(5) Leak monitor with shut-off valve and probe (Art.-No. 01056 – optional), recommended with fixed water
connection
(6) Connection for water level monitoring in the waste water container
(7) Waste water connection
(8) Connection of feed water
(9) Waste water container
(10) Water tap (available on-site)
(11) T-piece with hose (Art.-No. 38605)
(12) Feed water filter MELA
dem
(13) Cooling water outlet fitting
(14) Reverse-osmosis unit MELA
dem
47
(15) Double hose nozzle with anti-flooding flap (Art. No. 37400)
(16) Double-chamber trap (Art. No. 26635)
(17) Water tap DN40 (available on-site)
(18) Electrical connection