Chapter 9 – Malfunctions
52
Warning message
Possible cause
What you can do
Output medium is not ready The steam sterilizer is operated
without output medium although an
output medium is registered
Select the option in the menu
Batch
output
No output medium
(see page
35,
Initialisation of the log printer
).
Output medium is not correctly
connected and/or CF card is not
inserted
Check if the CF card of the CF card printer
is correctly inserted.
Check the correct connection of the data
cable to the steam sterilizer and at the
output medium.
Electric power supply of the log
printer or the CF card printer is
interrupted
Log printer is "offline"
Ensure electric power supply, the red LED
"P" at the log printer MELA
42/44 must
light red.
Set printer to "online" (press key "SEL" at
the MELA
42/44, LED "SEL" must light
green).
Protocol memory full
The device-internal log memory is
allocated (capacity got up to 40
logs).
The message is shown at the start of a
program.
Repeatedly press the K
EY
(S) to delete the
message, and the program starts. The
oldest log is then deleted.
An output medium is registered and
in the menu
Batch output
the
option
Immediate output NO
is
set.
Set printer to
Immediate output YES
(see page 38,
Automatic immediate log
output
).
Delete internal log memory (see page 40 ,
Delete saved logs
In orde
preset
deleted
In the f
), possibly output all saved logs beforehand
(see page 39,
Output all saved logs
).
Log off the output medium in the menu
batch output output medium
and
set the option
no output medium
.
Execute Service
The servicing message is activated,
the device has reached the
prescribed batch number
The message is shown after every program
start.
Repeatedly press the K
EY
(S) to delete the
message, and the program starts.
Retain the message: press the (S) K
EY
twice
to start.
Have maintenance carried out by MELAG
customer service or service of the specialist
dealer.
PLEASE NOTE
Service counter is reset by customer
service.
Warning/Battery empty
Monitoring of the device's internal
battery voltage determined a too
low voltage value.
The battery must be exchanged by MELAG
customer service or service of the specialist
dealer.