9.4 No display on the screen
______________________________________________
After switching on the autoclave, the display should show the initial setting (see section 8).
If there is no display, check:
1. Is the cable plugged into the mains?
2. Is the mains supply O.K:(if necessary check with another appliance)?
3. Check the three mains fuses. These fuses are installed behind the door. Since the electric door
opening function may in such a case also no longer function, then please proceed as described
under section 9.5.
Disconnect the power cable and remove the screw caps over the fuses using a screwdriver or a
coin. Exchange the fuses (three reserve fuses are delivered with the autoclave) then replace the
screw cap and reconnect the autoclave to the power supply. If there is still no display when the
autoclave is switched on, or if the display blacks out repeatedly, please inform your specialist dealer.
If you exchange the fuses, order two new spare fuses through your dealer (MELAG Art.-No. 57590).
9.5 Emergency measure if the electric door release button fails
_________________
In the unlikely event that the electric door system should fail: In such a case, you can use the crank handle
delivered with the autoclave to open the door as follows:
12.1
Remove the cap (12) from the emergency door-opening socket.
12.2
Insert the crank handle into the socket that is now open.
To
open
the door, turn the crank to the left (counter-clockwise, or anti-clockwise): see Fig. 7 below.
To
close
the door, turn the crank to the right (clockwise).
13
12
1
2
12.1
12.2
Fig. 7: Emergency operation of the door latch
WARNING:
If you use the crank handle to open the door, it is very important to pull out the crank
BEFORE
the door is completely open. If you open the door completely before pulling out the crank handle, this will
damage the plastic door.