User Manual for the Cliniklav
25
20
4. This is followed by ventilation of the
autoclave chamber.
After the ventilation of the chamber the
display shows that the program has been
successfully completed …
… in alternation with "Last batch number
4 - Quit with '+' ".
If a printer or another output medium is
connected and the option "Immediate
output" is set to "YES" a log with
confirmation that the drying process has
been terminated is printed out.
By pressing the key '+' the door is
unlocked …
… and can be opened after the message
"Open door please" is displayed.
4.11 Removing the sterilized items
__________________________________________
Warning
Be careful when removing the sterilized items!
Touching the metal surfaces can lead to
burns
. Always use the appropriate aids to lift the trays or wear suitable hand protection.
4.12 Sterile storage
______________________________________________________
After removing wrapped sterile items, the wrapping should be checked for any signs of damage. If it is
defective (e.g. split seals) then the sterilization of the items must be repeated after the items have been re-
wrapped.
It is important for sterile storage that the items have been properly dried. The Cliniklav
®
25 provides very
good drying if the program has not been interrupted before its completion and the autoclave has been
properly loaded (see Section 4.3). Directly after sterilization there may still be residual condensation on the
items or the container. Because the items are hot on removal, this will usually evaporate quickly. The
German industrial standard DIN 58953 Part 7 Section 7 contains the following comment about residual
moisture on paper wrapping or transparent sterilization paper after sterilization: "...small amounts of water on
the wrapping are unproblematic, provided they have evaporated within 30 minutes after removal from the
steam sterilizer....."
After cooling, wrapped sterilized objects should be stored in a place where they are
protected from dust
(e.g. instrument cupboard). Given proper storage, the sterilized objects can be stored up to 6 months
according to DIN .8953-8:2003.
Program Start-Stop
O p e n d o or
p l ea se
Program Start-Stop
U n i ver sal -Pr og r am
r u n s uc ce ss f ul l y
Program Start-Stop
V e n ti l a ti o n
- 0. 52 b ar 11 2° C
Program Start-Stop
D r yi n g
s t o p pe d
Program Start-Stop
L a s t ba tc h n um b er 4
Q ui t w i th '+ '