User Manual for the Cliniklav
25
41
The internal testing of the sensors for temperature,
pressure or conductivity showed an excessive deviation,
the malfunction can be reported on switching on the
autoclave or in the course of a program:
•
If this occurs repeatedly, inform your specialist dealer.
The limit deviation for internal analogue/digital signal
conversion has been exceeded:
•
If this occurs repeatedly, inform your specialist dealer.
The limit deviation between the two sensors for the steam
temperature has been exceeded:
•
If this occurs repeatedly, inform your specialist dealer.
During the Vacuum test program the pressure was too high
(very large leak):
•
Repeat the vacuum test, and if there is another
malfunction message inform your specialist dealer.
After starting the program there was a loss of power. The
error report is received when the electricity supply is
restored:
•
Check the mains power supply installation, if no errors
can be found, inform the service agent.
If there is a loss of power when the chamber is under
pressure, then there will be an additional reminder to
sterilize the sterile filter, since this may have become moist
and non-sterile:
•
Remove the sterile filter at the rear of the autoclave.
•
Sterilize the filter using the Quick-Program S.
•
Then replace the filter.
The time limit for the steam generator to reach the
necessary pressure has been exceeded:
•
If this occurs repeatedly, inform your specialist dealer.
The minimum sterilization temperature has not been
reached:
•
Reduce the size of the load.
•
If this occurs repeatedly, inform your specialist dealer.
The maximum sterilization temperature has been
exceeded:
•
If this occurs repeatedly, inform your specialist dealer.
Sterilization pressure falls below the minimum level:
•
Reduce the size of the load.
•
If this occurs repeatedly, inform your specialist dealer.
The maximum sterilization pressure has been exceeded:
•
If this occurs repeatedly, inform your specialist dealer.
The difference between measured and theoretical
temperature is too large:
•
If this occurs repeatedly, inform your specialist dealer.
M al f u nc ti o n 32 :
P ow er f ai l ur e
S t er i l e fi l ter
s t er i l i z e
M al f u nc ti o n 18 :
S e n so r : . . . I npu t : . . .
M al f u nc ti o n 26 :
A/ D - C o n ve r t er
M al f u nc ti o n 27 :
T e m p. S e ns . d ef 1, 2
M al f u nc ti o n 31 :
S ys t e m l ea k
M al f u nc ti o n 33 :
P r e ss. r el ea se
M al f u nc ti o n 34 :
S t er i l i z a ti o n TU 1
M al f u nc ti o n 35 :
S t er i l i z a ti o n TO 1
M al f u nc ti o n 36 :
S t er i l i z a ti o n D U
M al f u nc ti o n 37 :
S t er i l i z a ti o n D O
M al f u nc ti o n 38 :
S t er i l i z a ti o n TD 1