„Duration“ determines the duration of a measurement. The sam-
pling frequency resulting from "Samples" and "Duration" is displayed.
2.2.3.9 Amplitude
The “Amplitude” menu opens a window for entering the measuring
range for both channels.
The amplitude can be set between 0.2 V and 20 V in the grid 1-2-
5. The selection values adapt to the set division factor of the
probe. The default value is 20 V.
2.2.3.10 Probe
The “Probe” menu opens a window for entering voltage dividers for
signal conditioning.
If voltages higher than ± 10 V are to be measured, a probe with a
division factor of 10 or 100 must be used. This factor should be in-
dicated next to the amplitude. All voltage-related information is
based on this factor. Please note that most probes with sub-fac-
tors greater than “1x” are not very accurate.