the last 512 bytes. This corresponds to 256 or 128 measuring
points, depending on the mode. The modes Oscilloscope and Logic-
Analyzer create a file only partially described or with no content at
all, in case of early termination. These files do not have time
stamps and their content is undefined. They should be deleted prior
to the evaluation of other measurement data.
1.6 System Requirements
System Requirements for MEphisto Scope UM202/UM203 und
MEphistoLab
2
: Windows 7 through Windows 10 as well as driver
software ME-iDS:
Please note the corresponding README-files.
1.7 Trigger Modes
Manual (Manual)
This trigger-mode stands for the freely running measurement.
It starts immediately after pressing „Run“ or „Single“.
Threshold value (Level, above / Level, below)
In this mode, the oscilloscope reacts to exceeding or not meet-
ing the indicated threshold. It is not important how the ex-
ceeded threshold value was created. This trigger mode is espe-
cially suitable for two cases: