Model FL-PWJ Power Supply & FL-LHJ 
Lamphouse

 

 

1. Introduction 

 

The Meiji Techno FL-PWJ Epi-Fluorescence Illuminator Power Supply is a 
precision engineered product incorporating efficient, silent-running switchmode 
power supply technology delivering smooth, regulated current. 
 
The FL-PWJ Power Supply is designed only to support 100 watt HBO DC arc 
lamps. 
 
The advanced switchmode design provides constant, low-noise DC power that 
automatically adjusts for the condition of the DC lamp in circuit. 
 
The FL-PWJ Power Supply comes standard with a panel mounted circuit breaker 
and provides accidental unplug protection to guard against laboratory mishaps. 
 
A panel mounted Elapsed Time Panel Meter keeps track of the DC arc lamp 
hours in use so replacements can be easily accommodated. 
 
The reset button on the panel meter resets to “zero” when arc lamps are 
replaced. 
 
This system was designed for easy plug-n-play integration with 100 watt HBO 
illumination commonly used in fluorescence microscopy work with no additional 
adjustments necessary for the lifespan of the arc lamp. 
 
The FL-PWJ Power Supply can be used with the TC Series Inverted Epi-
Fluorescence Microscope and comes standard on the MT6000 Series Epi-
Fluorescence Upright Biological Models. 

 
 
 
 
 
 

Содержание FL-LHJ

Страница 1: ...Model FL PWJ Power Supply Model FL LHJ Lamphouse Instruction Manual...

Страница 2: ...Information 1 6 Warranty Notes 2 0 The Components 2 1 Installation Site 2 2 Unpacking 2 3 Set Up 2 4 Bulb Installation or Replacement Step by Step 2 5 Bulb Alignment for TC Models 2 6 Bulb Alignment...

Страница 3: ...standard with a panel mounted circuit breaker and provides accidental unplug protection to guard against laboratory mishaps A panel mounted Elapsed Time Panel Meter keeps track of the DC arc lamp hour...

Страница 4: ...J Power Supply This manual should be read carefully before any attempt is made to operate this equipment To ensure safe operation the user must read and adhere to all of the directions put forth in th...

Страница 5: ...use only as a biological microscope illumination system Modifying this product in any way for use in any situation other than the original and intended product design will automatically void the warra...

Страница 6: ...se during or after operation Do not touch the lamp house or attempt to replace the bulb for at least 30 minutes after the unit has been turned off or injury may result Do not obstruct the air vents on...

Страница 7: ...d by abuse neglect misuse improper operation normal wear accident improper maintenance or modifications of ANY type This warranty does not cover repair or replacement where normal use has exhausted th...

Страница 8: ...for merchantability or fitness for a particular purpose or otherwise Liability for consequential damages under any and all warranties are excluded to the extent exclusions are permitted by law This wa...

Страница 9: ...connector 6 IEC Standard inlet connector for standard IEC powercord sets 7 DC Arc Lamp Access Panel Screws 2 8 X Y adjustment for integrated concave mirror 9 Concave Mirror Position Set Screw 10 Conc...

Страница 10: ...80 up to 30 C No dew condensation allowed 2 2 Unpacking Please check your packing slip to insure that all materials are present Keep a copy for your records so that you have the proper information whe...

Страница 11: ...until they stop and gently removing the access plate If the DC Arc lamp is not installed proceed to install the DC arc lamp in the correct orientation shown below being careful not to touch the glass...

Страница 12: ...from the cable to the panel connector When properly aligned the cable connector twists and engages the panel connector and screws all the way inwards The power supply is ready to be plugged into the w...

Страница 13: ...and cause pre mature failure If you have any difficulty inserting a new lamp into the socket it may be possible that the stamping lettering on the bulb could be too raised To insure a good contact the...

Страница 14: ...knobs clockwise to lock the door 9 Plug the socket into the power supply and plug the power supply into the outlet 10 Ignite lamp by turning on the power supply and adjust lamp house controls to get...

Страница 15: ...any methods 1 If the arc lamp is not already burning turn on the power supply and let the arc lamp settle for at least 5 minutes 2 On the stage of the TC place a piece of plastic on the stage that flu...

Страница 16: ...o achieve the brightest spot in the center of the FOV or field of view 5 While still looking into the eyetubes turn the knob on the lamphouse which controls the focus of the collector lens assembly No...

Страница 17: ...seen the FOV you do not have the field iris centered AND the field iris not open all the way 2 6 Bulb Alignment MT Models After the bulb has burned in the system is ready for arc lamp alignment There...

Страница 18: ...on the lamphouse and adjust both the X and the Y controls to achieve the brightest spot in the center of the FOV or field of view 5 While still looking into the eyetubes turn the knob on the lamphouse...

Страница 19: ...minor adjustments to the arc position and the mirror position until the FOV displays an even an image as possible 9 If a shadow is seen the FOV you do not have the field iris centered AND the field i...

Страница 20: ...longer performing accurately 3 1 UV Light Safety Considerations Mercury arc lamps have intense UV light content in their output The U S National Institute for Occupational Safety and Health NIOSH rec...

Страница 21: ...rations should be taken very seriously Limit access to areas where UV light is present Post warning signs in the area where the equipment is installed Always wear protective eyewear and gloves Be sure...

Страница 22: ...ping your microscope clean can help the overall quality of your work Cleaning of Painted Surfaces Use a soft brush or lint free cotton cloth to removed dust and loose particles Tough dirt can be remov...

Страница 23: ...in regard to specimens fluorochrome fluorescence specimen preparation specimen staining etc are beyond the scope of this manual 5 1 Fluorescence Lamp does not work LAMPHOUSE MAY BE HOT TO THE TOUCH D...

Страница 24: ...ex but there is a followed procedure that will make the process easier and less dangerous If the glass envelope of the mercury bulb happens to break open you and the immediate environment could be exp...

Страница 25: ...Output Stability Ripple 2 5 P P Starting Voltage Peak 2 3KV DC Operating Temperature 0 40 degrees C Operating Humidity 20 85 RH Storage Temperature 20 70 degrees C Storage Humidity 10 90 RH Supplied A...

Страница 26: ...ttp www meijitechno com MEIJI TECHNO CO LTD 322 1 Chikumazawa Miyoshi machi Iruma gun Saitama 354 0043 Japan Phone Fax E mail Web 049 259 0111 049 259 0113 meiji meijitechno co jp http www meijit echn...