3. Slide desired lamp into the lamp holder.
4. Slide the bulb clamp onto the negative (-) end of the lamp.
5. Hold the bulb clamp and trigger lead. Using the supplied allen wrench, tighten
the clamping screws snugly and evenly.
Be careful not to place ANY stress
on the lamp.
6. Tighten two set screws on (+) side of lamp.
7. Pick up side panel with mounted lamp. Note markings on side panel. Rotate
panel until appropriate end has the text and indexer aligned correctly.
8. Insert panel in lamp house opening. Rotate and tighten the two black knobs
clockwise to lock the door.
9. Plug the socket into the power supply and plug the power supply into the
outlet.
10. Ignite lamp by turning on the power supply and adjust lamp house controls to
get proper arc alignment within the field of view.
Refer to the microscope instructions for more information regarding arc
alignment. Once alignment has been set, it will not need any further adjustment
until a new bulb is installed.
If you have any problems with this procedure, please contact your microscope
sales person.
PROPER SAFETY GOGGLES AND PROTECTION ARE
NECCESARY WHEN OPERATING EQUIPMENT THAT HAS ULTRA-VIOLET
LIGHT CONTENT.
11. A new bulb should ALWAYS be run for two hours initially to stabilize the arc
of the lamp.