9.
Calibration Kit Manager
•
Connector definition
•
Cal Kits and priority
•
Cal Kit Element definition
The VNA software has a Cal Kit Manager that keeps track of available and required
calibration elements and their coefficients. The software needs to know what kind of
connectors are used and their gender so that it can apply the correct compensation
coefficients for each calibration.
On the left is a list of available connector types. These are the types that you can select
for each measurement port. You can add new types or remove them, unless a connector
type is in use by one of the calibration kits. If a calibration kit is imported with a new
connector type this type will be added to the list automatically.
The Cal Kits panel shows the available calibration kit types. You can add new kits or
remove them, or import a kit from a Cal Kit file (.vck). You can also export a Cal Kit.
The Copy key can be used to copy the selected Cal Kit to a new one, to make it easy to
start a new Cal Kit.
With the arrows you can move a Cal Kit up or down in the list. The list position provides
a priority in case there are multiple Cal Kits with the same connector. The software will
assign a calibration element from the highest Cal Kit in the list.