Depending on the measurement
one or more of the port panes are
enabled. For each port select the
Connector Type
and
Gender
.
The software will automatically
search the available calibration
kits for suitable and matching
calibration elements.
Connectors can have a
Female
or
a
Male
gender. The
Neutral
gender is used for calibration kits
without a gender, like PCB or test
fixture calibration kits.
These are the
Genders
of the
Measurement Connectors
.
When one or more calibration
elements are not found in the kits,
the screen can be closed but the
setup is not complete and no
calibrations can be made.
The
Cal Kits
button opens the
Calibration Kit Manager
screen. This feature is described elsewhere in this manual.
Sweep
The VNA can sweep one or more parameters to make a measurement. The most common
sweep is the simple Frequency sweep, with the frequency going from the start to stop
value. All other parameters stay at a fixed value during the measurement.
Other useful sweeps include stepping the measurement power for linearity measurements
or stepping the bias voltage or current to measure active elements.
The sweep tab shows the VNA parameters and which of them are involved in a sweep
(blue arrow) and which are fixed value (blue vertical bar).