![MegiQ VNA04 Series Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/megiq/vna04-series/vna04-series_user-manual_1765660026.webp)
VNA0440
–
VNA0440e
VNA0460
–
VNA0460e
user manual V3.1
- 26 -
•
Connect the measuring cable and press the
Sweep
button to perform an impedance
measurement of the open circuit.
•
In the
Display tab
, select the appropriate S-parameter
and press the
Set Open
button.
•
The reference plane has now moved to the position of
the open end and the following measurements will be at
that point.
It is also possible to perform this procedure with
shorting
the measuring point to ground
and use the
Set Short
button.
With a
two-port
measurement a through measurement can be used to deembed the
through
path.