6
SYNC
light (
23
):
When unit is synchronized to the power line source, this lamp will be lit. The line
synchronization is selected from the
Timer Setup Screen
, see
Out Freq.
description.
INIT
(INITIATE) Button (
24
):
The
INIT
Button serves to initiate (turn on) the
CURRENT MAIN
output of the test set.
In addition, it also serves to initiate certain test functions like a timing test on voltage
relay.
PRINT SCRN
Button (
25
):
This button is used in conjunction with the Parallel Printer Port to print the metered / test
values from the display screen.
Up, Down, Right and Left Arrows
(
26
):
These arrows move the graphic screen highlighted function up, down, right and left on
the screen. It works in conjunction with the rotary knob to change the highlighted
function.
INPUT POWER
There are two versions of the SR-98. One is rated for 230 volt operation only. The
other has a universal input of 90 to 253 volts, 1700 VA max.
Model
SR-98-1/60
comes with a universal power input (90 - 253 VAC, 50/60 Hz input),
a north American power cord and firmware set to default to 60 Hz output on power up.
Model
SR-98-2/50
comes with a universal power input (90 - 253 VAC, 50/60 Hz input),
a international color coded power cord and firmware set to default to 50 Hz output on
power up. Model
SR-98-3/50
has a power input of
230 volt only
(230 VAC, 50/60 Hz
input), a Continental Europe power cord and firmware set to default to 50 Hz output on
power up.