18
METERING Screen:
The METERING Screen allows the SR-98 to measure external voltage and current, as a
multi-purpose single-phase meter. AC or DC voltages up 600 Volts and AC or DC
currents up to 6 Amperes may be measured. Other measured and calculated values
are displayed such as Phase Angle, VAR, Power and Power Factor.
Function buttons allow the user to select either Amperes AC or DC and Volts AC or DC.
VOLTAGE RELAY Test Screen:
The VOLTAGE RELAY Test Screen is used to test all single-phase AC under and
overvoltage relays.
The
SET FAULT VOLTS
button is used to preset the desired “fault” under or over
voltage value.
AC VOLTS ( 00.0 V ):
The value displayed is the measured output voltage from the
AC AUX output terminals. The voltage value displayed in the ( 00.0 V ) is the estimated
output for the selected output terminal. The actual output voltage may vary slightly
depending on the actual load. This allows the operator to preset a voltage output
without turning the voltage on. When rotating the Control Knob, the voltage value
changes up or down depending on the direction of rotation.