24
SERVICE DATA
MAINTENANCE INSTRUCTIONS
Maintenance intervals depend on usage, but a maximum of every six months is
recommended.
WARNING:
Do not service unit unless it is disconnected from its power source.
1.
Enclosure: The enclosure can be cleaned with a soft cloth. If heavily soiled, the
cloth can be dampened with an approved solvent that does not attack the finish
or leave residue.
2.
Control Panel: The control panel can be wiped clean with a soft, dry cloth. Do
not wipe the meter lens with a cloth. If a breath of air will not remove dirt, brush it
away lightly with a soft-bristle instrument brush.
3.
Other Components: Check all knobs, printed circuit boards, screws, fasteners,
connections and terminals for tightness and proper position. Remove dust with a
soft brush and breath of air. Output terminal connection tightness is particularly
important. If they become loose, excessive heating of the terminals and poor
current output will result.
4.
Insulation: Check wiring and other insulated components for burning, cracking or
other damage.
IMPORTANT NOTES
Do not use lubricants or solvents of any kind in the test set except as specifically
recommended.
If damage or malfunction is suspected or repairs deemed necessary, consult a
MEGGER Representative for assistance if it is unclear what course of action is needed.
Be sure to provide all name plate data when making inquiries.
BASIC TROUBLESHOOTING
The troubleshooting information relies on the technician to have a through
understanding of the operation of the unit. If the technician is unfamiliar with the unit, he
or she should not attempt to repair. The technician should contact the factory before
attempting repairs. Provide the Megger part number for the part or assembly in
question and the serial number of the SR-98 when making inquiries.