To select Mode (Method)
Scroll to
“Mode”
(also known as method),
to configure the integrated MTDR100.
Select the required mode by rotating the
job-dial until “Mode” is highlighted. Then
by depressing the jog-dial, you will then be
presented with the following sub-menu
options:
Arc Reflection:
Use to select Arc
Reflection and Arc Reflection plus method
of pre-location on the MTDR
TDR:
To select standard Pulse Echo
method.
Current Impulse:
Select for Current
Impulse or Impulse Current
Voltage Decay:
To select Voltage Decay
method (not available on PFL22M1500)
Additional menu options in TDR mode
Once the
TDR mode
has been selected the operator menu bar will change,
additional items only highlighted.
Traces:
Allows the operator to select which
traces he wants to view, and which traces
are active
Show Traces
: the operator can choose
between the tree traces which are viewed.
Tick to activate. You are then able to
compare any of the three phases. By using
the vertical control (see later) you can
overlay each of the traces to look for any
deviations
Содержание PFL22M1500 Series
Страница 2: ......
Страница 16: ...M AVTMPFL22M Rev 1 April 2011 6 M...
Страница 28: ...M AVTMPFL22M Rev 1 April 2011 18 M...
Страница 40: ...M AVTMPFL22M Rev 1 April 2011 30 M...
Страница 43: ...AVTMPFL22M Rev 1 April 2011 33 Addendum 1 MTDR100 User Guide...
Страница 44: ...M AVTMPFL22M Rev 1 April 2011 34 M...
Страница 61: ...AVTMPFL22M Rev 1 April 2011 51 Addendum 2 PFL22M1500INV XX PFL22M1500INV Inverter Option only...
Страница 62: ...M AVTMPFL22M Rev 1 April 2011 52 M...
Страница 68: ...M AVTMPFL22M Rev 1 April 2011 58 M...
Страница 69: ...AVTMPFL22M Rev 1 April 2011 59 Addendum 3 Cable Fault Location Applications Guide...
Страница 70: ...M AVTMPFL22M Rev 1 April 2011 60...
Страница 77: ...AVTMPFL22M Rev 1 April 2011 67 Megger Quality System Certificate...
Страница 78: ...M AVTMPFL22M Rev 1 April 2011 68...