45
Specification
Circuit test
Test Voltage:
12 V p - p, 50Hz square wave. (Typical)
Max current:
100 mA (typical)
Indications:
Open circuit, OK, Short circuit.
Insulation test
Range:
0 - 50 M
Ω
Accuracy:
5.0% ± 100 k
Ω
Compliance:
500 V at 0.5 M
Ω
Pass Band options:
0.25M
Ω
, 0.3M
Ω
, 0.5M
Ω
, 1M
Ω
, 2M
Ω
, 4M
Ω
, 5M
Ω
, 7M
Ω
, 8M
Ω
, 10M
Ω
, 20M
Ω
, 50M
Ω
Earth Bond Test
Range:
0 - 999 m
Ω
Resolution 1 m
Ω
1000 – 1999m
Ω
Resolution 10 m
Ω
Accuracy:
0 <R> 499 m
Ω
5.0% ±5m
Ω
500 m <R> 1999 m
Ω
5.0% ±50m
Ω
Test Voltage:
10 V a.c. (Typical)
Test Current:
10A or 25A into 0.1 (typical)
Pass Bands:
50 m
Ω
, 100 m
Ω
, 200 m
Ω
, 300 m
Ω
, 400m
Ω
, 500m
Ω
, 600m
Ω
, 750m
Ω
, 800m
Ω
, 1
Ω
,
1.5, 2
Earth Continuity Test
Range:
0 - 999 m
Ω
Resolution 1 m
Ω
1.00 - 9.99
Ω
Resolution 10 m
Ω
Accuracy:
5.0% ±10m
Ω
Test Voltage:
4 V d.c. (Typical)
Test Current:
200 mA short circuit (typical)
Pass Bands:
50 m
Ω
, 100 m
Ω
, 200 m
Ω
, 500 m
Ω
, 750m
Ω
, 1
Ω
, 1.5
Ω
, 2
Ω
, 2.5
Ω
, 5
Ω
, 7.5
Ω
, 10
Ω
Operation Test
Range:
0 - 3.0 kVA.
Accuracy:
0 <VA> 99 VA:
5.0% ± 5 VA
100 <VA> 999 VA:
5.0% ± 10 VA
1.0 <VA> 3.0kVA:
5.0% ± 100 VA
Reference:
Reading corrected to 230 V
Pass Band options:
50 VA, 100 VA, 200 VA, 350 VA, 500 VA, 700 VA, 1000 VA, 1200 VA, 1500VA,
2000 VA, 2500 VA, 3000 VA.
Содержание PAT4DV/3
Страница 1: ...User Guide M PAT4DV 3 4DVF 3 Portable Appliance Testers...
Страница 49: ...49 Repair and Warranty...
Страница 50: ...50 Notes...
Страница 51: ...M Part Number 6172 706 V13 Printed in England 1108 www megger com...