Each test can be performed for a User definable time of between 5 and 60 seconds (5 and 20 seconds for Bond test). This enables
specific tests to be set to appropriate lengths of time, perhaps to satisfy a particular standard. Once set, a test time will automatically
be applied to all assets using such a test, regardless of the Group, Client or Location.
1 From the Main Screen, press the
MENU
key.
2 Press the
SETTINGS
key.
3 Press the
MISC
key.
4 Press the
TEST SETUP
key.
5 Press the
TEST TIMES
key. The first test is displayed with its current time setting.
6 Operate the left hand keys to adjust the time for the selected test, then press the
OK
key to
store the test time as displayed. The next test and it’s current duration is displayed.
Note: Press the
ESC
ape key at any time to exit test time setup.
7 When the final test time has been setup and
OK
pressed, the display returns to the Test
Setup screen.
Setup Test Times
Flash test
5 s
Operation test
30 s
Insulation test
5 s
Continuity test
5 s
Bond test
5 s