When the test has stopped, the display continues to show the voltage 
present on the test leads. Pressing either test voltage 

s

 or 

t

 button will 

then display the test voltage immediately before the end of the test.

Test start / stop button

A test will only start if this button is pressed, held and then released as 
soon as the red high voltage warning indicator LED lights. The LCD and 
red LED high voltage warning indicators flash when the test starts.

A test will not start if the button is released before the red LED shows, or 
pressed continuously for longer than 5 seconds. This is a safety feature to 
prevent a test being started inadvertently.

The presence of a voltage greater than 50 V on the test leads is indicated 
with flashing high voltage warning indicators. Testing is disabled if the 
external voltage exceeds 80 V.

Testing will stop if the test start / stop button is pressed again, the 
preset test time is reached, or, if the unit is not in burn mode, insulation 
breakdown is detected.

When a test has finished, the instrument will discharge the load, which 
may take some time. The operator must always check the load has been 
discharged before touching the test leads.

Ω

/I button

Pressing this button toggles the digital and secondary displays. The details 
available when scrolling the secondary display will depend upon the test 
mode selected, whether timers T1 and T2 have been set, and the duration 
of the test. The digital display toggles between insulation resistance and 
current.

Power On/Off button

The instrument will only turn on if this button is pressed, held and then 
released when the display responds. The instrument will not turn on if 
the button is released before the display responds, or if the button is held 
down for too long. This is a safety feature to prevent the instrument being 
inadvertently turned on.

The instrument is turned off either by pressing the button again, or if the 
instrument is running on the battery, by timing out after 10 minutes of 
inactivity.

Upon switching the instrument on the display will first show ‘Ini’ while it 
undergoes a self-checking routine. When Ini disappears, the instrument is 
ready for use. 

Test voltage 

s

 

and

 

t

 

buttons

Using these buttons one of six test voltages can be selected: 250 V, 500 V,  
1 kV, 2.5 kV, 5 kV, and 10 kV. The selected voltage is shown on the display.

A non standard voltage between 50 V and 10 kV can be selected by holding 
down the ‘Fn’ function button whilst operating the buttons. The selectable 
voltage is adjustable in 10 V steps between 50 V and 1 kV, and adjustable in 
25 V steps between 1 kV and 10 kV. An auto repeat facility is enabled when 
the button is held down, allowing faster travel through the range.

If there is an external voltage greater than 50 V on the test leads, the high 
voltage warning indicators are flashed, and the display shows this voltage 
instead. The instrument will not perform a test if this voltage is greater than 
80 V.

During a test the display shows the actual voltage on the test leads. If the 
test voltage is changed during a test, the new test voltage will be displayed 
briefly. 

Содержание MIT1020/2

Страница 1: ...ess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demo...

Страница 2: ...M MIT1020 2 10kV Digital Insulation Tester USER MANUAL...

Страница 3: ...manner possibly causing a loss of display while the circuit remains energised In this event the unit must be turned off and the circuit discharged manually n Test leads including crocodile clips must...

Страница 4: ...circumstances Failure to follow these instructions could result in damage to the instrument and or a hazard to the operator The word should is used to indicate that the instructions indicate best prac...

Страница 5: ...instrument performs automatic tests and has data storage and data retrieval facilities Polarization index and dielectric discharge tests are performed automatically and test duration and voltages can...

Страница 6: ...on table Connection UK International USA Earth Ground Yellow Green Green Neutral Blue White Live Line Brown Black If using a fused plug ensure that it is fitted with a 3 Amp fuse The instrument can be...

Страница 7: ...dicator Alarm Data recording Test modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main T1 T2 Dielectr...

Страница 8: ...es between insulation resistance and current Power On Off button The instrument will only turn on if this button is pressed held and then released when the display responds The instrument will not tur...

Страница 9: ...ry display initially shows the PI polarity index DAR dielectric absorption ratio and on completion of the test the TC time constant and capacitance measurements Toggling the display shows insulation r...

Страница 10: ...ndition test runs for longer than 10 minutes as this is required for a PI reading Figure 2 The result of a polarity index test PI Test settings T1 and T2 times set in order to measure the DAR Test con...

Страница 11: ...the test mode Modes of test to be chosen from include an insulation resistance IR test a polarization index PI test a step voltage SV test and a dielectric discharge DD test The insulation resistance...

Страница 12: ...instrument now displays clr to indicate clear mode Press and hold down test button until clear process begins shown by analogue display lighting anti clockwise Press the MODE button to exit without d...

Страница 13: ...terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if the insulator is clean and th...

Страница 14: ...lt to that selected in the previous test If the instrument has been switched off since the previous test the test voltage defaults to 250 V If there is an external voltage greater than 80 V present th...

Страница 15: ...voltage in test voltage in test voltage in test voltage in major steps major steps minor steps minor steps IR only IR only Vs AND Vt Sets voltage Show set to 500V voltage Ts Tt Increase Display set ti...

Страница 16: ...ments and can provide accurate measurements with up to 4mA of noise current present A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the...

Страница 17: ...ent symbol PI On completion of the test the instrument will display the insulation resistance recorded at 1 minute and 10 minutes under the segment symbols 1m and 16 TEST MODE SUMMARY 10m Pressing the...

Страница 18: ...limits of 10 k and 35 T Release the function button to save the current limit Adjusting the alarm threshold level automatically activates the alarm limit mode This is indicated by the A symbol turning...

Страница 19: ...leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 G the test leads should not be allowed to touc...

Страница 20: ...ads are available as an optional accessory from Megger The lead to the negative terminal is fully screened The screen is plugged into the Guard terminal diverting any stray leakage currents This consi...

Страница 21: ...t accuracy 23 C 5 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V of nominal test voltage at 1 G load Current measurement range 0 01 nA to 5 mA Current measurement accuracy 23 C 5 0 2 nA at all volt...

Страница 22: ...ips 1002 534 Optional Accessories HV test lead sets 3 x 3 m with un insulated small clips 8101 181 3 x 8 m with un insulated small clips 8101 182 3 x 15 m with un insulated small clips 8101 183 3 x 5...

Страница 23: ...PA 19403 England U S A Tel 44 0 1304 502 243 Tel 1 610 676 8579 Fax 44 0 1304 207 342 Fax 1 610 676 8625 Megger operate fully traceable calibration and repair facilities ensuring your instrument conti...

Страница 24: ...ttery charge level indicator will need to calibrated to the new battery s characteristics End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispo...

Страница 25: ...723 2861 USA ONLY T 1 214 333 3201 F 1 214 331 7399 E ussales megger com Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES France T 33 0 1 30 16 08 90 F 33 0 1 34 61 23 77 E infos...

Страница 26: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: