background image

15

TEST MODE SUMMARY

On screen warning codes

Breakdown detection

 

If insulation breakdown is detected, 
the unit aborts the test and flashes 
“brd” and the final test result 
alternately in the main digital display. 
This display is disabled when in burn 
mode as the unit will continue the 
test without aborting. The instrument 
detects breakdown in terms of a rapid 
drop in test voltage commensurate 
with that caused by an arc across or though the insulation under test.

Noise detection

Excessive noise during test

In environments such as HV switch yards or substations electrical noise 
levels can be high enough to effect insulation test measurements. This 
noise takes the form of induced currents in the measurement circuit. 

The noise current is either picked up by test leads or induced directly 
in the item under test. The MIT510/2, MIT520/2 and MIT1020/2 are all 
capable of accurate measurements with up to 2mA of induced noise 
current. The S1-554/2 and S1-1054/4 are design for very high noise 
environments and can provide accurate measurements with up to 4mA of 
noise current present.

A level of noise exceeding the instruments specified capability will be 
indicated by an alternating sine wave on the display. The test continues 
while displaying the noise. If the noise drops back down the normal 
display is resumed.

The effects of noise can be reduces significantly with the use of high 
quality Megger screened test leads.

The picture displayed on the right 
shows the expected screen result 
of an IR test on with a 5000V test 
voltage after 2 minutes 9 seconds. 
The black and grey waveforms 
are displayed alternately while the 
noise is present. When the noise is 
removed the wave form is replaced 
by the correct reading.

Test aborted due to extremely high noise

Should an extremely high level of noise 
be present, particularly at the start of 
a test, the instrument will indicate nSE 
(Noise). The instrument will require 
switching off and on again to reset.

In certain rare circumstances 
breakdown of the circuit under test 
may interfere with the sensitive current 
measuring circuit. In this event the unit stops and also displays nSE. The 
instrument will again require switching off and on again to reset.

Содержание MIT1020/2

Страница 1: ...ess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demo...

Страница 2: ...M MIT1020 2 10kV Digital Insulation Tester USER MANUAL...

Страница 3: ...manner possibly causing a loss of display while the circuit remains energised In this event the unit must be turned off and the circuit discharged manually n Test leads including crocodile clips must...

Страница 4: ...circumstances Failure to follow these instructions could result in damage to the instrument and or a hazard to the operator The word should is used to indicate that the instructions indicate best prac...

Страница 5: ...instrument performs automatic tests and has data storage and data retrieval facilities Polarization index and dielectric discharge tests are performed automatically and test duration and voltages can...

Страница 6: ...on table Connection UK International USA Earth Ground Yellow Green Green Neutral Blue White Live Line Brown Black If using a fused plug ensure that it is fitted with a 3 Amp fuse The instrument can be...

Страница 7: ...dicator Alarm Data recording Test modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main T1 T2 Dielectr...

Страница 8: ...es between insulation resistance and current Power On Off button The instrument will only turn on if this button is pressed held and then released when the display responds The instrument will not tur...

Страница 9: ...ry display initially shows the PI polarity index DAR dielectric absorption ratio and on completion of the test the TC time constant and capacitance measurements Toggling the display shows insulation r...

Страница 10: ...ndition test runs for longer than 10 minutes as this is required for a PI reading Figure 2 The result of a polarity index test PI Test settings T1 and T2 times set in order to measure the DAR Test con...

Страница 11: ...the test mode Modes of test to be chosen from include an insulation resistance IR test a polarization index PI test a step voltage SV test and a dielectric discharge DD test The insulation resistance...

Страница 12: ...instrument now displays clr to indicate clear mode Press and hold down test button until clear process begins shown by analogue display lighting anti clockwise Press the MODE button to exit without d...

Страница 13: ...terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if the insulator is clean and th...

Страница 14: ...lt to that selected in the previous test If the instrument has been switched off since the previous test the test voltage defaults to 250 V If there is an external voltage greater than 80 V present th...

Страница 15: ...voltage in test voltage in test voltage in test voltage in major steps major steps minor steps minor steps IR only IR only Vs AND Vt Sets voltage Show set to 500V voltage Ts Tt Increase Display set ti...

Страница 16: ...ments and can provide accurate measurements with up to 4mA of noise current present A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the...

Страница 17: ...ent symbol PI On completion of the test the instrument will display the insulation resistance recorded at 1 minute and 10 minutes under the segment symbols 1m and 16 TEST MODE SUMMARY 10m Pressing the...

Страница 18: ...limits of 10 k and 35 T Release the function button to save the current limit Adjusting the alarm threshold level automatically activates the alarm limit mode This is indicated by the A symbol turning...

Страница 19: ...leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 G the test leads should not be allowed to touc...

Страница 20: ...ads are available as an optional accessory from Megger The lead to the negative terminal is fully screened The screen is plugged into the Guard terminal diverting any stray leakage currents This consi...

Страница 21: ...t accuracy 23 C 5 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V of nominal test voltage at 1 G load Current measurement range 0 01 nA to 5 mA Current measurement accuracy 23 C 5 0 2 nA at all volt...

Страница 22: ...ips 1002 534 Optional Accessories HV test lead sets 3 x 3 m with un insulated small clips 8101 181 3 x 8 m with un insulated small clips 8101 182 3 x 15 m with un insulated small clips 8101 183 3 x 5...

Страница 23: ...PA 19403 England U S A Tel 44 0 1304 502 243 Tel 1 610 676 8579 Fax 44 0 1304 207 342 Fax 1 610 676 8625 Megger operate fully traceable calibration and repair facilities ensuring your instrument conti...

Страница 24: ...ttery charge level indicator will need to calibrated to the new battery s characteristics End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispo...

Страница 25: ...723 2861 USA ONLY T 1 214 333 3201 F 1 214 331 7399 E ussales megger com Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES France T 33 0 1 30 16 08 90 F 33 0 1 34 61 23 77 E infos...

Страница 26: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: