minute period. Readings for contaminated insulation are fairly constant 
because any absorption effects are masked by high leakage currents. 
On completion of the test the polarisation index is displayed under the 
segment heading ‘PI’.

Polarisation Index (PI) = R

10 min

                                       R

1 min 

The ‘Dielectric Absorption Ratio’ is the term applied to the polarisation 
index using other time intervals set by T1 and T2.

If timers T1 and T2 are both set, then the insulation resistance measured 
at these times is also recorded. These are displayed together with the 
calculated dielectric absorption ratio under the T1, T2, and DAR segments 
respectively.

The instrument can be made to display either the insulation resistances, 
insulation currents or ratios plus capacitance measurements by using the 

Ω

/I toggle button. See section ‘

Ω

/I button’ for cycling through results and 

toggling units.

Dielectric Absorption Ratio (DAR) = R

T2 min

  

 

 

       R

T1 min

Dielectric discharge ‘DD’ test

The ‘DD’ test is a diagnostic insulation test that allows ageing, 
deterioration, and voids in the insulation to be assessed. The result is 
dependent on the discharge characteristic, so the internal condition of the 
insulation is tested, largely independent of any surface contamination. On 
discharge the capacitive component of the discharge current decays from 

17

the ‘A 

­

’ symbol is flashing on the display. Use the timer 

s

 and 

t

 buttons 

to set the resistance threshold between the limits of 10 k

Ω

 and 35 T

Ω

Release the function button to save the current limit. Adjusting the alarm 
threshold level automatically activates the alarm limit mode. This is 
indicated by the ‘A 

­

’ symbol turning solidly on.

To toggle the alarm limit mode on/off, hold down the function button, 
then press and release the mode button repeatedly until the ‘A 

­

’ symbol is 

flashing on the display. Release the function key to toggle.

Step voltage ‘SV’ test

This is a test based on the principle that an ideal insulator will produce 
identical readings at all voltages, while an insulator which is being over 
stressed, will show lower insulation values at higher voltages. The main 
timer and test voltage settings can be adjusted if desired from their default 
values of 5 minutes and 1 kV respectively. During the test the applied test 
voltage incrementally steps by one fifth of the test voltage setting final 
value every minute, for 5 minutes, taking successive measurements until 
the final voltage is reached. Readings for the first 4 recorded values are 
displayed under the consecutive segment headings ‘1m’ to ‘4m’. The 5 
minute reading is displayed by the main display.

If the range of measured insulation between readings is too wide for the 
instrument to display, then those readings too small compared to the final 
reading will be represented by  ‘---’. 

Polarisation index ‘PI’ test, and dielectric absorption ratio 
‘DAR’ test

The ‘PI’ test is a particular example of a time/resistance method, which 
takes the ratio of the insulation measured at 1 minute and at 10 minutes. 
Good insulation generally shows an increase in resistance over a 10 

Содержание MIT1020/2

Страница 1: ...ess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demo...

Страница 2: ...M MIT1020 2 10kV Digital Insulation Tester USER MANUAL...

Страница 3: ...manner possibly causing a loss of display while the circuit remains energised In this event the unit must be turned off and the circuit discharged manually n Test leads including crocodile clips must...

Страница 4: ...circumstances Failure to follow these instructions could result in damage to the instrument and or a hazard to the operator The word should is used to indicate that the instructions indicate best prac...

Страница 5: ...instrument performs automatic tests and has data storage and data retrieval facilities Polarization index and dielectric discharge tests are performed automatically and test duration and voltages can...

Страница 6: ...on table Connection UK International USA Earth Ground Yellow Green Green Neutral Blue White Live Line Brown Black If using a fused plug ensure that it is fitted with a 3 Amp fuse The instrument can be...

Страница 7: ...dicator Alarm Data recording Test modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main T1 T2 Dielectr...

Страница 8: ...es between insulation resistance and current Power On Off button The instrument will only turn on if this button is pressed held and then released when the display responds The instrument will not tur...

Страница 9: ...ry display initially shows the PI polarity index DAR dielectric absorption ratio and on completion of the test the TC time constant and capacitance measurements Toggling the display shows insulation r...

Страница 10: ...ndition test runs for longer than 10 minutes as this is required for a PI reading Figure 2 The result of a polarity index test PI Test settings T1 and T2 times set in order to measure the DAR Test con...

Страница 11: ...the test mode Modes of test to be chosen from include an insulation resistance IR test a polarization index PI test a step voltage SV test and a dielectric discharge DD test The insulation resistance...

Страница 12: ...instrument now displays clr to indicate clear mode Press and hold down test button until clear process begins shown by analogue display lighting anti clockwise Press the MODE button to exit without d...

Страница 13: ...terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if the insulator is clean and th...

Страница 14: ...lt to that selected in the previous test If the instrument has been switched off since the previous test the test voltage defaults to 250 V If there is an external voltage greater than 80 V present th...

Страница 15: ...voltage in test voltage in test voltage in test voltage in major steps major steps minor steps minor steps IR only IR only Vs AND Vt Sets voltage Show set to 500V voltage Ts Tt Increase Display set ti...

Страница 16: ...ments and can provide accurate measurements with up to 4mA of noise current present A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the...

Страница 17: ...ent symbol PI On completion of the test the instrument will display the insulation resistance recorded at 1 minute and 10 minutes under the segment symbols 1m and 16 TEST MODE SUMMARY 10m Pressing the...

Страница 18: ...limits of 10 k and 35 T Release the function button to save the current limit Adjusting the alarm threshold level automatically activates the alarm limit mode This is indicated by the A symbol turning...

Страница 19: ...leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 G the test leads should not be allowed to touc...

Страница 20: ...ads are available as an optional accessory from Megger The lead to the negative terminal is fully screened The screen is plugged into the Guard terminal diverting any stray leakage currents This consi...

Страница 21: ...t accuracy 23 C 5 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V of nominal test voltage at 1 G load Current measurement range 0 01 nA to 5 mA Current measurement accuracy 23 C 5 0 2 nA at all volt...

Страница 22: ...ips 1002 534 Optional Accessories HV test lead sets 3 x 3 m with un insulated small clips 8101 181 3 x 8 m with un insulated small clips 8101 182 3 x 15 m with un insulated small clips 8101 183 3 x 5...

Страница 23: ...PA 19403 England U S A Tel 44 0 1304 502 243 Tel 1 610 676 8579 Fax 44 0 1304 207 342 Fax 1 610 676 8625 Megger operate fully traceable calibration and repair facilities ensuring your instrument conti...

Страница 24: ...ttery charge level indicator will need to calibrated to the new battery s characteristics End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispo...

Страница 25: ...723 2861 USA ONLY T 1 214 333 3201 F 1 214 331 7399 E ussales megger com Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES France T 33 0 1 30 16 08 90 F 33 0 1 34 61 23 77 E infos...

Страница 26: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: