www.megger.com
17
Operation
Making a continuity measurement
1. Set the Primary (Left) range knob to
W
range. (The position of the right hand rotary range knob must not be
in the
position.)
2. Connect two test leads to the L1 (+ve) and L2 (-ve) terminals on the instrument. A continuity measurement is
made automatically.
L
N
E
g
NOTE:
Measurements are prevented when:
NOTE:
A resistance of > 99.9 kΩ is present
NOTE:
Circuit voltages in excess of 4 V are detected.
Storing / downloading results
Once the display shows a value it will automatically be logged into temporary memory. Unless stored, this will be
over written by the next measurement.
To store this result or to send it to a compatible device, refer to “Sending, Storing, Deleting and Recalling Test
Results” on page 29.
Continuity Buzzer ON/OFF
Whilst in the continuity range, press the MODE button
. This will toggle the buzzer ON and OFF.
Buzzer ON =
Buzzer OFF = No symbol