18
Electrical Specification
Accuracy is expressed as:
±
(% of r no. of least significant
digits) at 23
±
5
and <75% R.H.
Resistance (parallel mode)
Test Frequency: 120Hz / 1KHz
Accuracy
Range
Maximum
Display
@120 Hz
@1KHz
Specified Note
10M
9.999M
2.0%+8
*3
2.0%+8
*3
After
open
cal.
2000K
1999.9K
0.5%+5 0.5%+5
After
open
cal.
200K
199.99K
0.5%+3 0.5%+3
-
20K
19.999K
0.5%+3 0.5%+3
-
2000
1999.9
0.5%+3 0.5%+3
-
200
199.99
0.8%+5 0.8%+5
After
short
cal.
20
19.999
1.2%+40 1.2%+40
After
short
cal.
Notes:
1. This specification is based on measurements performed at the test
socket.
2. DUT (Device Under Test) & Test leads should be properly shielded
to GUARD if necessary.
3. This specification is based on battery operation.
Capacitance (parallel mode)
Test Frequency: 120 Hz
Accuracy
Range Maximum
Display
Capacitance
DF
Spec. Note
10mF 19.99mF
*5
3.0%+5
(DF<0.1)
10%+100/Cx+5
(DF<0.1)
After short
cal.
1000
μ
F 1999.9
μ
F
*6
1.0%+5
(DF<0.1)
2%+100/Cx+5
(DF<0.1)
After short
cal.
200
μ
F 199.99
μ
F
0.7%+3
DF<0.5
0.7%+100/Cx+5
(DF<0.5)
-
20
μ
F 19.999
μ
F
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
2000nF 1999.9nF 0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
200nF 199.99nF 0.7%+5
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
After open
cal.
20nF 19.999nF 1.0%+5
(DF<0.1)
2%+100/Cx+5
(DF<0.1)
After open
cal.