11
Operating Instructions
Data Hold
This data hold function allows the operator to freeze the
display. To enter this mode, press the "
HOLD
" pushbutton;
press again to release.
Static Recording
Press the "
REC
" pushbutton for more than one second to
enter the static recording mode. The maximum and
minimum readings are then stored in memory, while a
beeping tone is produced when a new tested value has been
recorded. Push the same button to cycle through the current
maximum, minimum and average readings.
The
MAX
,
MIN
or
AVG
indicators on LCD will turn on to
indicate which value is shown.
To exit this mode, press and hold the pushbutton for more
than one second.
Notes:
1. Static recording captures only stable values and updates
the memory; it will not record any "
OL
" (overload) value for
any of the
L/C/R
functions. In addition, the meter will not
record values below 50 counts in Capacitance
measurement.
2. Static recording is only available in manual ranging;
however, activation while in auto-ranging will automatically
set meter to manual ranging and cause calibration prompts
to be displayed in the relevant ranges.
Dissipation Factor / Quality Factor
The "
D/Q
" value can be displayed alternately by pressing the
"
D/Q
" button when the meter is set to Inductance or
Capacitance mode. It does not apply to resistance
measurement.