19
Test Frequency: 1 KHz
Accuracy
Range
Maximum
Display
Capacitance
DF
Spec. Note
1mF 1.999mF
*5
3.0%+5
(DF<0.1)
10%+100/Cx+5
(DF<0.1)
After short
cal.
200
μ
F 199.99
μ
F
1.0%+5
(DF<0.1)
2.0%+100/Cx+5
(DF<0.1)
After short
cal.
20
μ
F 19.999
μ
F
0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
2000nF 1999.9nF 0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
200nF 199.99nF 0.7%+3
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
-
20nF 19.999nF 0.7%+5
(DF<0.5)
0.7%+100/Cx+5
(DF<0.5)
After open
cal.
2000pF 1999.9pF 1.0%+5
(DF<0.1)
2.0%+100/Cx+5
(DF<0.1)
After open
cal.
Notes:
1. Q Value is the reciprocal of DF.
2. This specification is based on measurements performed at
the test socket.
3. DUT & Test leads should be properly shielded to GUARD
if necessary.
4. Cx=Counts of displayed C value, e.g. if C=88.88
μ
F, then
Cx=8888.
5. This reading can be extended to 1999 MAX display with
accuracy not specified.
6. This reading can be extended to 19999 MAX display with
accuracy not specified.